{"_buckets": {"deposit": "3981b1fb-45e5-4c2f-9ec9-47cb04c137d3"}, "_deposit": {"created_by": 3, "id": "16517", "owners": [3], "pid": {"revision_id": 0, "type": "depid", "value": "16517"}, "status": "published"}, "_oai": {"id": "oai:kanazawa-u.repo.nii.ac.jp:00016517", "sets": ["1259"]}, "author_link": ["30349", "20089", "30350"], "item_7_alternative_title_2": {"attribute_name": "その他のタイトル", "attribute_value_mlt": [{"subitem_alternative_title": "乳房撮影における吸収物質内の線質について"}]}, "item_7_biblio_info_8": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "1996-12-18", "bibliographicIssueDateType": "Issued"}, "bibliographicPageEnd": "48", "bibliographicPageStart": "43", "bibliographicVolumeNumber": "20", "bibliographic_titles": [{"bibliographic_title": "金沢大学医学部保健学科紀要 = Memoirs of School of Health Sciences Faculty of Medicine Kanazawa University"}]}]}, "item_7_description_21": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "In order to evaluate the x-ray quality in each material, we obtained the values from one to five half-value layer with aluminum, bakelite, polymethyl methacrylate resin and BR-12 phantom, and investigated the change of x-ray quality (from first to fifth half-value layer) according to the attenuation in their materials. The un-homogeneity factor to first half-value layer in BR-12 phantom was slightly low than that in aluminum, and especially was not as high as that in aluminum for the ratio of fifth half-value layer to first half-value layer. X-ray quality passed through the breast may be higher than that of the breast surface because of the transmission in the breast tissue. So the subject contrast would be decreasing. The thickness of the breast under compression in the clinical practice correspond to about four or five half-value layer. Therefore the x-ray quality must be evaluated by BR-12 phantom, being composed nearly with the breast.", "subitem_description_type": "Abstract"}]}, "item_7_publisher_17": {"attribute_name": "出版者", "attribute_value_mlt": [{"subitem_publisher": "金沢大学医学部保健学科 = School of Health Sciences Faculty of Medicine Kanazawa University"}]}, "item_7_source_id_11": {"attribute_name": "NCID", "attribute_value_mlt": [{"subitem_source_identifier": "AA11137625", "subitem_source_identifier_type": "NCID"}]}, "item_7_source_id_9": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "1342-7318", "subitem_source_identifier_type": "ISSN"}]}, "item_7_version_type_25": {"attribute_name": "著者版フラグ", "attribute_value_mlt": [{"subitem_version_resource": "http://purl.org/coar/version/c_970fb48d4fbd8a85", "subitem_version_type": "VoR"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Koshida, Kichiro"}], "nameIdentifiers": [{"nameIdentifier": "20089", "nameIdentifierScheme": "WEKO"}, {"nameIdentifier": "90020023", "nameIdentifierScheme": "e-Rad", "nameIdentifierURI": "https://kaken.nii.ac.jp/ja/search/?qm=90020023"}, {"nameIdentifier": "90020023", "nameIdentifierScheme": "研究者番号", "nameIdentifierURI": "https://nrid.nii.ac.jp/nrid/1000090020023"}]}, {"creatorNames": [{"creatorName": "Mizushima, Takashi"}], "nameIdentifiers": [{"nameIdentifier": "30349", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Yamamoto, Tomoyuki"}], "nameIdentifiers": [{"nameIdentifier": "30350", "nameIdentifierScheme": "WEKO"}]}]}, "item_files": {"attribute_name": "ファイル情報", "attribute_type": "file", "attribute_value_mlt": [{"accessrole": "open_date", "date": [{"dateType": "Available", "dateValue": "2017-10-04"}], "displaytype": "detail", "download_preview_message": "", "file_order": 0, "filename": "AA11137625-20-43-48.pdf", "filesize": [{"value": "448.2 kB"}], "format": "application/pdf", "future_date_message": "", "is_thumbnail": false, "licensetype": "license_free", "mimetype": "application/pdf", "size": 448200.0, "url": {"label": "AA11137625-20-43-48.pdf", "url": "https://kanazawa-u.repo.nii.ac.jp/record/16517/files/AA11137625-20-43-48.pdf"}, "version_id": "5929159a-fd0c-440e-bdbb-4db053ce83e5"}]}, "item_keyword": {"attribute_name": "キーワード", "attribute_value_mlt": [{"subitem_subject": "X-ray quality", "subitem_subject_scheme": "Other"}, {"subitem_subject": "Half-value layer", "subitem_subject_scheme": "Other"}, {"subitem_subject": "Subject contrast", "subitem_subject_scheme": "Other"}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "eng"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "departmental bulletin paper", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "Study of X-ray Quality in Absorption Materials for Mammography", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "Study of X-ray Quality in Absorption Materials for Mammography"}]}, "item_type_id": "7", "owner": "3", "path": ["1259"], "permalink_uri": "http://hdl.handle.net/2297/46400", "pubdate": {"attribute_name": "公開日", "attribute_value": "2017-10-04"}, "publish_date": "2017-10-04", "publish_status": "0", "recid": "16517", "relation": {}, "relation_version_is_last": true, "title": ["Study of X-ray Quality in Absorption Materials for Mammography"], "weko_shared_id": 3}
Study of X-ray Quality in Absorption Materials for Mammography
http://hdl.handle.net/2297/46400
http://hdl.handle.net/2297/46400c65127fe-ff60-4db2-89df-ebc343eaa685
名前 / ファイル | ライセンス | アクション |
---|---|---|
AA11137625-20-43-48.pdf (448.2 kB)
|
|
Item type | 紀要論文 / Departmental Bulletin Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-10-04 | |||||
タイトル | ||||||
タイトル | Study of X-ray Quality in Absorption Materials for Mammography | |||||
言語 | ||||||
言語 | eng | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | X-ray quality | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | Half-value layer | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | Subject contrast | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | departmental bulletin paper | |||||
その他のタイトル | ||||||
その他のタイトル | 乳房撮影における吸収物質内の線質について | |||||
著者 |
Koshida, Kichiro
× Koshida, Kichiro× Mizushima, Takashi× Yamamoto, Tomoyuki |
|||||
書誌情報 |
金沢大学医学部保健学科紀要 = Memoirs of School of Health Sciences Faculty of Medicine Kanazawa University 巻 20, p. 43-48, 発行日 1996-12-18 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 1342-7318 | |||||
NCID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AA11137625 | |||||
出版者 | ||||||
出版者 | 金沢大学医学部保健学科 = School of Health Sciences Faculty of Medicine Kanazawa University | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | In order to evaluate the x-ray quality in each material, we obtained the values from one to five half-value layer with aluminum, bakelite, polymethyl methacrylate resin and BR-12 phantom, and investigated the change of x-ray quality (from first to fifth half-value layer) according to the attenuation in their materials. The un-homogeneity factor to first half-value layer in BR-12 phantom was slightly low than that in aluminum, and especially was not as high as that in aluminum for the ratio of fifth half-value layer to first half-value layer. X-ray quality passed through the breast may be higher than that of the breast surface because of the transmission in the breast tissue. So the subject contrast would be decreasing. The thickness of the breast under compression in the clinical practice correspond to about four or five half-value layer. Therefore the x-ray quality must be evaluated by BR-12 phantom, being composed nearly with the breast. | |||||