{"_buckets": {"deposit": "e4571bf0-4b9b-4b9e-9822-9fbeb04dd5bf"}, "_deposit": {"created_by": 3, "id": "14606", "owners": [3], "pid": {"revision_id": 0, "type": "depid", "value": "14606"}, "status": "published"}, "_oai": {"id": "oai:kanazawa-u.repo.nii.ac.jp:00014606", "sets": ["1134"]}, "author_link": ["25866", "1059", "25865"], "item_4_biblio_info_8": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2011-06-01", "bibliographicIssueDateType": "Issued"}, "bibliographicIssueNumber": "2", "bibliographicPageEnd": "76", "bibliographicPageStart": "72", "bibliographicVolumeNumber": "39", "bibliographic_titles": [{"bibliographic_title": "Journal of Nuclear Medicine Technology"}]}]}, "item_4_description_21": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "The choice of injected dose of 18F-FDG and acquisition time is important in obtaining consistently high-quality PET images. The aim of this study was to determine the optimal acquisition protocols based on patient weight for 3-dimensional lutetium oxyorthosilicate PET/CT. Methods: This study was a retrospective analysis of 76 patients ranging from 29 to 101 kg who were injected with 228-395.2 MBq of 18F-FDG for PET imaging. The study population was divided into 4 weightbased groups: less than 45 kg (group 1), 45-59 kg (group 2), 60-74 kg (group 3), and 75 kg or more (group 4). We measured the true coincidence rate, random coincidence rate, noise-equivalent counting rate (NECR), and random fraction and evaluated image quality by the coefficient of variance (COV) in the largest liver slices. Results: The true coincidence rate, random coincidence rate, and NECR significantly increased with increasing injected dose per kilogram (r 5 0.91, 0.83, and 0.90; all P \u003c 0.01). NECR maximized at 10.11 MB/kg in underweight patients. The true coincidence rate differed significantly among the 4 groups, except for group 3 versus group 4 (P \u003c 0.01). The ratio of the true coincidence rate for group 2 to groups 3 and 4 was 1.4 and 1.6, respectively. The average random fraction for all 4 groups was approximately 35%. The COV of the 4 groups differed for all pairs (P \u003c 0.01). The COVs in overweight patients were larger than those in underweight patients, and image quality in overweight patients was poor. Conclusion: We modified acquisition protocols for 18F-FDG PET/CT according to the characteristics of a 3-dimensional lutetium orthosilicate PET scanner and PET image quality based on patient weight. The optimal acquisition time was approximately 1.4-1.6 times longer in overweight patients than in normal-weight patients. Estimation of optimal acquisition times using the true coincidence rate is more important than other variables in improving PET image quality. © 2011 by the Society of Nuclear Medicine, Inc.", "subitem_description_type": "Abstract"}]}, "item_4_description_22": {"attribute_name": "内容記述", "attribute_value_mlt": [{"subitem_description": "Thesis of Nagaki, Akio / 長木 昭男 博士学位論文(金沢大学 / 大学院医薬保健学総合研究科)", "subitem_description_type": "Other"}]}, "item_4_publisher_17": {"attribute_name": "出版者", "attribute_value_mlt": [{"subitem_publisher": "Society of Nuclear Medicine"}]}, "item_4_relation_12": {"attribute_name": "DOI", "attribute_value_mlt": [{"subitem_relation_type": "isVersionOf", "subitem_relation_type_id": {"subitem_relation_type_id_text": "10.2967/jnmt.110.081661", "subitem_relation_type_select": "DOI"}}]}, "item_4_source_id_11": {"attribute_name": "NCID", "attribute_value_mlt": [{"subitem_source_identifier": "AA00703709", "subitem_source_identifier_type": "NCID"}]}, "item_4_source_id_9": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "0091-4916", "subitem_source_identifier_type": "ISSN"}]}, "item_4_version_type_25": {"attribute_name": "著者版フラグ", "attribute_value_mlt": [{"subitem_version_resource": "http://purl.org/coar/version/c_ab4af688f83e57aa", "subitem_version_type": "AM"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Nagaki, Akio"}], "nameIdentifiers": [{"nameIdentifier": "25865", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Onoguchi, Masahisa"}], "nameIdentifiers": [{"nameIdentifier": "1059", "nameIdentifierScheme": "WEKO"}, {"nameIdentifier": "30283102", "nameIdentifierScheme": "研究者番号", "nameIdentifierURI": "https://nrid.nii.ac.jp/nrid/1000030283102"}]}, {"creatorNames": [{"creatorName": "Matsutomo, Norikazu"}], "nameIdentifiers": [{"nameIdentifier": "25866", "nameIdentifierScheme": "WEKO"}]}]}, "item_files": {"attribute_name": "ファイル情報", "attribute_type": "file", "attribute_value_mlt": [{"accessrole": "open_date", "date": [{"dateType": "Available", "dateValue": "2017-10-03"}], "displaytype": "detail", "download_preview_message": "", "file_order": 0, "filename": "ME-PR-NAGAKI-A-72.pdf", "filesize": [{"value": "486.5 kB"}], "format": "application/pdf", "future_date_message": "", "is_thumbnail": false, "licensetype": "license_free", "mimetype": "application/pdf", "size": 486500.0, "url": {"label": "ME-PR-NAGAKI-A-72.pdf", "url": "https://kanazawa-u.repo.nii.ac.jp/record/14606/files/ME-PR-NAGAKI-A-72.pdf"}, "version_id": "c4edc954-158b-4bcf-9259-793c8ac51119"}]}, "item_keyword": {"attribute_name": "キーワード", "attribute_value_mlt": [{"subitem_subject": "18F-FDG", "subitem_subject_scheme": "Other"}, {"subitem_subject": "Acquisition protocols", "subitem_subject_scheme": "Other"}, {"subitem_subject": "Image quality", "subitem_subject_scheme": "Other"}, {"subitem_subject": "Patient weight", "subitem_subject_scheme": "Other"}, {"subitem_subject": "PET/CT", "subitem_subject_scheme": "Other"}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "eng"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "journal article", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "Patient weight-based acquisition protocols to optimize18F-FDG PET/CT image quality", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "Patient weight-based acquisition protocols to optimize18F-FDG PET/CT image quality"}]}, "item_type_id": "4", "owner": "3", "path": ["1134"], "permalink_uri": "http://hdl.handle.net/2297/30543", "pubdate": {"attribute_name": "公開日", "attribute_value": "2017-10-03"}, "publish_date": "2017-10-03", "publish_status": "0", "recid": "14606", "relation": {}, "relation_version_is_last": true, "title": ["Patient weight-based acquisition protocols to optimize18F-FDG PET/CT image quality"], "weko_shared_id": -1}
Patient weight-based acquisition protocols to optimize18F-FDG PET/CT image quality
http://hdl.handle.net/2297/30543
http://hdl.handle.net/2297/305435693edd2-0d13-48b9-ac8e-f346bed61d4d
名前 / ファイル | ライセンス | アクション |
---|---|---|
ME-PR-NAGAKI-A-72.pdf (486.5 kB)
|
|
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-10-03 | |||||
タイトル | ||||||
タイトル | Patient weight-based acquisition protocols to optimize18F-FDG PET/CT image quality | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
著者 |
Nagaki, Akio
× Nagaki, Akio× Onoguchi, Masahisa× Matsutomo, Norikazu |
|||||
書誌情報 |
Journal of Nuclear Medicine Technology 巻 39, 号 2, p. 72-76, 発行日 2011-06-01 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0091-4916 | |||||
NCID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AA00703709 | |||||
DOI | ||||||
関連タイプ | isVersionOf | |||||
識別子タイプ | DOI | |||||
関連識別子 | 10.2967/jnmt.110.081661 | |||||
出版者 | ||||||
出版者 | Society of Nuclear Medicine | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | The choice of injected dose of 18F-FDG and acquisition time is important in obtaining consistently high-quality PET images. The aim of this study was to determine the optimal acquisition protocols based on patient weight for 3-dimensional lutetium oxyorthosilicate PET/CT. Methods: This study was a retrospective analysis of 76 patients ranging from 29 to 101 kg who were injected with 228-395.2 MBq of 18F-FDG for PET imaging. The study population was divided into 4 weightbased groups: less than 45 kg (group 1), 45-59 kg (group 2), 60-74 kg (group 3), and 75 kg or more (group 4). We measured the true coincidence rate, random coincidence rate, noise-equivalent counting rate (NECR), and random fraction and evaluated image quality by the coefficient of variance (COV) in the largest liver slices. Results: The true coincidence rate, random coincidence rate, and NECR significantly increased with increasing injected dose per kilogram (r 5 0.91, 0.83, and 0.90; all P < 0.01). NECR maximized at 10.11 MB/kg in underweight patients. The true coincidence rate differed significantly among the 4 groups, except for group 3 versus group 4 (P < 0.01). The ratio of the true coincidence rate for group 2 to groups 3 and 4 was 1.4 and 1.6, respectively. The average random fraction for all 4 groups was approximately 35%. The COV of the 4 groups differed for all pairs (P < 0.01). The COVs in overweight patients were larger than those in underweight patients, and image quality in overweight patients was poor. Conclusion: We modified acquisition protocols for 18F-FDG PET/CT according to the characteristics of a 3-dimensional lutetium orthosilicate PET scanner and PET image quality based on patient weight. The optimal acquisition time was approximately 1.4-1.6 times longer in overweight patients than in normal-weight patients. Estimation of optimal acquisition times using the true coincidence rate is more important than other variables in improving PET image quality. © 2011 by the Society of Nuclear Medicine, Inc. | |||||
内容記述 | ||||||
内容記述タイプ | Other | |||||
内容記述 | Thesis of Nagaki, Akio / 長木 昭男 博士学位論文(金沢大学 / 大学院医薬保健学総合研究科) | |||||
著者版フラグ | ||||||
出版タイプ | AM | |||||
出版タイプResource | http://purl.org/coar/version/c_ab4af688f83e57aa |