WEKO3
インデックスリンク
アイテム
{"_buckets": {"deposit": "09ff666d-35fd-4aee-8828-d3b65f595de2"}, "_deposit": {"created_by": 3, "id": "16466", "owners": [3], "pid": {"revision_id": 0, "type": "depid", "value": "16466"}, "status": "published"}, "_oai": {"id": "oai:kanazawa-u.repo.nii.ac.jp:00016466", "sets": ["1257"]}, "author_link": ["20089"], "item_7_biblio_info_8": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "1998-12-25", "bibliographicIssueDateType": "Issued"}, "bibliographicPageEnd": "109", "bibliographicPageStart": "101", "bibliographicVolumeNumber": "22", "bibliographic_titles": [{"bibliographic_title": "金沢大学医学部保健学科紀要 = Memoirs of School of Health Sciences Faculty of Medicine Kanazawa University"}]}]}, "item_7_description_21": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "A fundamental examination of X-ray spectra analysis for the improvement of clinical technique and x-ray equipment was conducted. A comparison of X-ray spectrum of the water and acrylic resin showed that the X-ray transmission ratio of acrylic resin was high for low energy regions, and low for high energy regions. This constitutes a considerable problem when acrylic resin is used. When filters made of aluminum, copper, niobium and molybdenum were inserted, it was found that the attenuation ratios could be obtained with these filters as if the attenuation ratio was calculated by using mass attenuation coefficients of Hubbell without stripping correction. The addition of the Nb and Mo filters reduced in the photons of lower x-ray energy, and transmitted the photons of high-energy x-rays. The addition of Al filter tended to have the effect, but less evident, while the effect of the Cu filter was between that of Nb and Mo and of Al. The X-ray transmission ratio of the screen was also measured. For the range from 40 keV to 70keV, the orthochromatic screen should reduced the photons of x-ray, but for any other range the regular screen should reduced the photons.", "subitem_description_type": "Abstract"}]}, "item_7_publisher_17": {"attribute_name": "出版者", "attribute_value_mlt": [{"subitem_publisher": "金沢大学医学部保健学科 = School of Health Sciences, Faculty of Medicine, Kanazawa University"}]}, "item_7_source_id_11": {"attribute_name": "NCID", "attribute_value_mlt": [{"subitem_source_identifier": "AA11137625", "subitem_source_identifier_type": "NCID"}]}, "item_7_source_id_9": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "1342-7318", "subitem_source_identifier_type": "ISSN"}]}, "item_7_version_type_25": {"attribute_name": "著者版フラグ", "attribute_value_mlt": [{"subitem_version_resource": "http://purl.org/coar/version/c_970fb48d4fbd8a85", "subitem_version_type": "VoR"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "越田, 吉郎"}], "nameIdentifiers": [{"nameIdentifier": "20089", "nameIdentifierScheme": "WEKO"}, {"nameIdentifier": "90020023", "nameIdentifierScheme": "e-Rad", "nameIdentifierURI": "https://kaken.nii.ac.jp/ja/search/?qm=90020023"}, {"nameIdentifier": "90020023", "nameIdentifierScheme": "研究者番号", "nameIdentifierURI": "https://nrid.nii.ac.jp/nrid/1000090020023"}]}]}, "item_files": {"attribute_name": "ファイル情報", "attribute_type": "file", "attribute_value_mlt": [{"accessrole": "open_date", "date": [{"dateType": "Available", "dateValue": "2017-10-04"}], "displaytype": "detail", "download_preview_message": "", "file_order": 0, "filename": "AA1137625-22-101-109.pdf", "filesize": [{"value": "2.2 MB"}], "format": "application/pdf", "future_date_message": "", "is_thumbnail": false, "licensetype": "license_free", "mimetype": "application/pdf", "size": 2200000.0, "url": {"label": "AA1137625-22-101-109.pdf", "url": "https://kanazawa-u.repo.nii.ac.jp/record/16466/files/AA1137625-22-101-109.pdf"}, "version_id": "85983c9f-f13d-4c78-8e0b-e88cddc4201a"}]}, "item_keyword": {"attribute_name": "キーワード", "attribute_value_mlt": [{"subitem_subject": "X-ray spectra", "subitem_subject_scheme": "Other"}, {"subitem_subject": "added filter", "subitem_subject_scheme": "Other"}, {"subitem_subject": "K absorption edge", "subitem_subject_scheme": "Other"}, {"subitem_subject": "transmission ratio", "subitem_subject_scheme": "Other"}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "jpn"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "departmental bulletin paper", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "付加フィルタによって減弱するX線スペクトル解析のための基礎的検討", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "付加フィルタによって減弱するX線スペクトル解析のための基礎的検討"}, {"subitem_title": "A fundamental examination for X-ray spectra analysis attenuating by added filter", "subitem_title_language": "en"}]}, "item_type_id": "7", "owner": "3", "path": ["1257"], "permalink_uri": "http://hdl.handle.net/2297/43776", "pubdate": {"attribute_name": "公開日", "attribute_value": "2017-10-04"}, "publish_date": "2017-10-04", "publish_status": "0", "recid": "16466", "relation": {}, "relation_version_is_last": true, "title": ["付加フィルタによって減弱するX線スペクトル解析のための基礎的検討"], "weko_shared_id": -1}
付加フィルタによって減弱するX線スペクトル解析のための基礎的検討
http://hdl.handle.net/2297/43776
http://hdl.handle.net/2297/437768b2817b5-674f-4421-84cf-ada4b7ea68b6
名前 / ファイル | ライセンス | アクション |
---|---|---|
AA1137625-22-101-109.pdf (2.2 MB)
|
|
Item type | 紀要論文 / Departmental Bulletin Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-10-04 | |||||
タイトル | ||||||
タイトル | 付加フィルタによって減弱するX線スペクトル解析のための基礎的検討 | |||||
タイトル | ||||||
言語 | en | |||||
タイトル | A fundamental examination for X-ray spectra analysis attenuating by added filter | |||||
言語 | ||||||
言語 | jpn | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | X-ray spectra | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | added filter | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | K absorption edge | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | transmission ratio | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | departmental bulletin paper | |||||
著者 |
越田, 吉郎
× 越田, 吉郎 |
|||||
書誌情報 |
金沢大学医学部保健学科紀要 = Memoirs of School of Health Sciences Faculty of Medicine Kanazawa University 巻 22, p. 101-109, 発行日 1998-12-25 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 1342-7318 | |||||
NCID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AA11137625 | |||||
出版者 | ||||||
出版者 | 金沢大学医学部保健学科 = School of Health Sciences, Faculty of Medicine, Kanazawa University | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | A fundamental examination of X-ray spectra analysis for the improvement of clinical technique and x-ray equipment was conducted. A comparison of X-ray spectrum of the water and acrylic resin showed that the X-ray transmission ratio of acrylic resin was high for low energy regions, and low for high energy regions. This constitutes a considerable problem when acrylic resin is used. When filters made of aluminum, copper, niobium and molybdenum were inserted, it was found that the attenuation ratios could be obtained with these filters as if the attenuation ratio was calculated by using mass attenuation coefficients of Hubbell without stripping correction. The addition of the Nb and Mo filters reduced in the photons of lower x-ray energy, and transmitted the photons of high-energy x-rays. The addition of Al filter tended to have the effect, but less evident, while the effect of the Cu filter was between that of Nb and Mo and of Al. The X-ray transmission ratio of the screen was also measured. For the range from 40 keV to 70keV, the orthochromatic screen should reduced the photons of x-ray, but for any other range the regular screen should reduced the photons. | |||||
著者版フラグ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 |