{"_buckets": {"deposit": "61924337-7fd0-4316-b77f-20efa1554047"}, "_deposit": {"created_by": 18, "id": "46873", "owners": [18], "pid": {"revision_id": 0, "type": "depid", "value": "46873"}, "status": "published"}, "_oai": {"id": "oai:kanazawa-u.repo.nii.ac.jp:00046873", "sets": ["2817"]}, "author_link": ["21060", "78834"], "item_9_biblio_info_8": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2017-05-17", "bibliographicIssueDateType": "Issued"}, "bibliographicPageStart": "4p.", "bibliographicVolumeNumber": "2014-04-01 - 2017-03-31", "bibliographic_titles": [{"bibliographic_title": "平成28(2016)年度 科学研究費補助金 基盤研究(C) 研究成果報告書"}, {"bibliographic_title": "2016 Fiscal Year Final Research Report", "bibliographic_titleLang": "en"}]}]}, "item_9_creator_33": {"attribute_name": "著者別表示", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Onoguchi, Masahisa"}], "nameIdentifiers": [{"nameIdentifier": "21060", "nameIdentifierScheme": "WEKO"}, {"nameIdentifier": "30283120", "nameIdentifierScheme": "e-Rad", "nameIdentifierURI": "https://kaken.nii.ac.jp/ja/search/?qm=30283120"}]}]}, "item_9_description_21": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "外膜と内膜がそれぞれ任意に作動する新しい心臓動態ファントムを構築した。SPECT/CT装置を用い,ファントムの心筋層にTc-99m溶液,心室部に造影剤を入れて心機能評価の検証を行った。今回開発した本ファントムは,ストローク比を任意に変えることで異なる心筋壁の動きや左室容積が可変できる。本ファントムを用いることで多様な臨床条件(心室容積,駆出率,物理的現象)を想定した技術的評価が可能と考えられた。", "subitem_description_type": "Abstract"}, {"subitem_description": "The conventional dynamic cardiac phantom used in the field of nuclear medicine has a structure for which the size of the external side of the heartis fixed and only the inner side moves anteroposterior. Therefore, its usefulness in technical evaluation is limited. Hence, we developed a new dynamic cardiac phantom in which the outer and inner membranes freely move. Using a SPECT/CT system, we performed validation by filling the myocardial layer of the dynamic cardiac phantom with solution and the ventricle part with contrast medium. We evaluated myocardial wall motions of 3 segments (basal, mid, and apical) by setting the stroke ratios at 20:20 (ventricle-to-myocardial layer ratio). The myocardial wall motions at the stroke ratio of 20:20 were 7.50 ± 0.44, 11.15 ± 0.56, and 9.90 ± 0.24 mm in the basal, mid, and apical segments, respectively. These results suggest that the use of this phantom can realize technical evaluation that presumes various clinical conditions.", "subitem_description_type": "Abstract"}]}, "item_9_description_22": {"attribute_name": "内容記述", "attribute_value_mlt": [{"subitem_description": "研究課題/領域番号:26460725, 研究期間(年度):2014-04-01 - 2017-03-31", "subitem_description_type": "Other"}, {"subitem_description": "出典:「一体型SPECT/CT装置における心電図同期X線CT撮影の減弱補正法の確立」研究成果報告書 課題番号26460725\n(KAKEN:科学研究費助成事業データベース(国立情報学研究所)) \n(https://kaken.nii.ac.jp/report/KAKENHI-PROJECT-26460725/26460725seika/)を加工して作成", "subitem_description_type": "Other"}]}, "item_9_description_5": {"attribute_name": "提供者所属", "attribute_value_mlt": [{"subitem_description": "金沢大学医薬保健研究域保健学系", "subitem_description_type": "Other"}]}, "item_9_identifier_registration": {"attribute_name": "ID登録", "attribute_value_mlt": [{"subitem_identifier_reg_text": "10.24517/00053200", "subitem_identifier_reg_type": "JaLC"}]}, "item_9_relation_28": {"attribute_name": "関連URI", "attribute_value_mlt": [{"subitem_relation_name": [{"subitem_relation_name_text": "https://kaken.nii.ac.jp/search/?qm=30283120"}], "subitem_relation_type_id": {"subitem_relation_type_id_text": "https://kaken.nii.ac.jp/search/?qm=30283120", "subitem_relation_type_select": "URI"}}, {"subitem_relation_name": [{"subitem_relation_name_text": "https://kaken.nii.ac.jp/grant/KAKENHI-PROJECT-26460725/"}], "subitem_relation_type_id": {"subitem_relation_type_id_text": "https://kaken.nii.ac.jp/grant/KAKENHI-PROJECT-26460725/", "subitem_relation_type_select": "URI"}}, {"subitem_relation_name": [{"subitem_relation_name_text": "https://kaken.nii.ac.jp/report/KAKENHI-PROJECT-26460725/26460725seika/"}], "subitem_relation_type_id": {"subitem_relation_type_id_text": "https://kaken.nii.ac.jp/report/KAKENHI-PROJECT-26460725/26460725seika/", "subitem_relation_type_select": "URI"}}]}, "item_9_version_type_25": {"attribute_name": "著者版フラグ", "attribute_value_mlt": [{"subitem_version_resource": "http://purl.org/coar/version/c_ab4af688f83e57aa", "subitem_version_type": "AM"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "小野口, 昌久"}], "nameIdentifiers": [{"nameIdentifier": "78834", "nameIdentifierScheme": "WEKO"}, {"nameIdentifier": "30283120", "nameIdentifierScheme": "e-Rad", "nameIdentifierURI": "https://kaken.nii.ac.jp/ja/search/?qm=30283120"}]}]}, "item_files": {"attribute_name": "ファイル情報", "attribute_type": "file", "attribute_value_mlt": [{"accessrole": "open_date", "date": [{"dateType": "Available", "dateValue": "2019-02-14"}], "displaytype": "detail", "download_preview_message": "", "file_order": 0, "filename": "ME-PR-ONOGUCHI-M-kaken 2017-4p.pdf", "filesize": [{"value": "266.4 kB"}], "format": "application/pdf", "future_date_message": "", "is_thumbnail": false, "licensetype": "license_11", "mimetype": "application/pdf", "size": 266400.0, "url": {"label": "ME-PR-ONOGUCHI-M-kaken 2017-4p.pdf", "url": "https://kanazawa-u.repo.nii.ac.jp/record/46873/files/ME-PR-ONOGUCHI-M-kaken 2017-4p.pdf"}, "version_id": "bacd7f35-729c-4626-9c60-1ff4e044d4fb"}]}, "item_keyword": {"attribute_name": "キーワード", "attribute_value_mlt": [{"subitem_subject": "心筋動態ファントム", "subitem_subject_scheme": "Other"}, {"subitem_subject": "心電図同期SPECT", "subitem_subject_scheme": "Other"}, {"subitem_subject": "SPECT/CT装置", "subitem_subject_scheme": "Other"}, {"subitem_subject": "減弱補正", "subitem_subject_scheme": "Other"}, {"subitem_subject": "心機能", "subitem_subject_scheme": "Other"}, {"subitem_subject": "心筋血流", "subitem_subject_scheme": "Other"}, {"subitem_subject": "心電図同期CT撮影法", "subitem_subject_scheme": "Other"}, {"subitem_subject": "被ばく線量", "subitem_subject_scheme": "Other"}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "jpn"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "research report", "resourceuri": "http://purl.org/coar/resource_type/c_18ws"}]}, "item_title": "一体型SPECT/CT装置における心電図同期X線CT撮影の減弱補正法の確立", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "一体型SPECT/CT装置における心電図同期X線CT撮影の減弱補正法の確立"}, {"subitem_title": "Establishment of attenuation correction method using myocardial gated X-ray/CT imaging in integrated SPECT/CT system", "subitem_title_language": "en"}]}, "item_type_id": "9", "owner": "18", "path": ["2817"], "permalink_uri": "https://doi.org/10.24517/00053200", "pubdate": {"attribute_name": "公開日", "attribute_value": "2019-02-14"}, "publish_date": "2019-02-14", "publish_status": "0", "recid": "46873", "relation": {}, "relation_version_is_last": true, "title": ["一体型SPECT/CT装置における心電図同期X線CT撮影の減弱補正法の確立"], "weko_shared_id": -1}
一体型SPECT/CT装置における心電図同期X線CT撮影の減弱補正法の確立
https://doi.org/10.24517/00053200
https://doi.org/10.24517/00053200349df4f4-b686-4d2a-a1bc-d0c373d27332
名前 / ファイル | ライセンス | アクション |
---|---|---|
ME-PR-ONOGUCHI-M-kaken 2017-4p.pdf (266.4 kB)
|
Item type | 報告書 / Research Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2019-02-14 | |||||
タイトル | ||||||
タイトル | 一体型SPECT/CT装置における心電図同期X線CT撮影の減弱補正法の確立 | |||||
タイトル | ||||||
言語 | en | |||||
タイトル | Establishment of attenuation correction method using myocardial gated X-ray/CT imaging in integrated SPECT/CT system | |||||
言語 | ||||||
言語 | jpn | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_18ws | |||||
資源タイプ | research report | |||||
ID登録 | ||||||
ID登録 | 10.24517/00053200 | |||||
ID登録タイプ | JaLC | |||||
著者 |
小野口, 昌久
× 小野口, 昌久 |
|||||
著者別表示 |
Onoguchi, Masahisa
× Onoguchi, Masahisa |
|||||
提供者所属 | ||||||
内容記述タイプ | Other | |||||
内容記述 | 金沢大学医薬保健研究域保健学系 | |||||
書誌情報 |
平成28(2016)年度 科学研究費補助金 基盤研究(C) 研究成果報告書 en : 2016 Fiscal Year Final Research Report 巻 2014-04-01 - 2017-03-31, p. 4p., 発行日 2017-05-17 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | 外膜と内膜がそれぞれ任意に作動する新しい心臓動態ファントムを構築した。SPECT/CT装置を用い,ファントムの心筋層にTc-99m溶液,心室部に造影剤を入れて心機能評価の検証を行った。今回開発した本ファントムは,ストローク比を任意に変えることで異なる心筋壁の動きや左室容積が可変できる。本ファントムを用いることで多様な臨床条件(心室容積,駆出率,物理的現象)を想定した技術的評価が可能と考えられた。 | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | The conventional dynamic cardiac phantom used in the field of nuclear medicine has a structure for which the size of the external side of the heartis fixed and only the inner side moves anteroposterior. Therefore, its usefulness in technical evaluation is limited. Hence, we developed a new dynamic cardiac phantom in which the outer and inner membranes freely move. Using a SPECT/CT system, we performed validation by filling the myocardial layer of the dynamic cardiac phantom with solution and the ventricle part with contrast medium. We evaluated myocardial wall motions of 3 segments (basal, mid, and apical) by setting the stroke ratios at 20:20 (ventricle-to-myocardial layer ratio). The myocardial wall motions at the stroke ratio of 20:20 were 7.50 ± 0.44, 11.15 ± 0.56, and 9.90 ± 0.24 mm in the basal, mid, and apical segments, respectively. These results suggest that the use of this phantom can realize technical evaluation that presumes various clinical conditions. | |||||
内容記述 | ||||||
内容記述タイプ | Other | |||||
内容記述 | 研究課題/領域番号:26460725, 研究期間(年度):2014-04-01 - 2017-03-31 | |||||
内容記述 | ||||||
内容記述タイプ | Other | |||||
内容記述 | 出典:「一体型SPECT/CT装置における心電図同期X線CT撮影の減弱補正法の確立」研究成果報告書 課題番号26460725 (KAKEN:科学研究費助成事業データベース(国立情報学研究所)) (https://kaken.nii.ac.jp/report/KAKENHI-PROJECT-26460725/26460725seika/)を加工して作成 |
|||||
著者版フラグ | ||||||
出版タイプ | AM | |||||
出版タイプResource | http://purl.org/coar/version/c_ab4af688f83e57aa | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://kaken.nii.ac.jp/search/?qm=30283120 | |||||
関連名称 | https://kaken.nii.ac.jp/search/?qm=30283120 | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://kaken.nii.ac.jp/grant/KAKENHI-PROJECT-26460725/ | |||||
関連名称 | https://kaken.nii.ac.jp/grant/KAKENHI-PROJECT-26460725/ | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://kaken.nii.ac.jp/report/KAKENHI-PROJECT-26460725/26460725seika/ | |||||
関連名称 | https://kaken.nii.ac.jp/report/KAKENHI-PROJECT-26460725/26460725seika/ |