WEKO3
インデックスリンク
アイテム
Graphite Furnace Atomic Absorption Spectrometric Determination of Bismuth(III) after Coprecipitation with Hafnium Hydroxide
http://hdl.handle.net/2297/38205
http://hdl.handle.net/2297/382059bb865ae-b8e3-4d7c-8b41-0136a1d7901c
| 名前 / ファイル | ライセンス | アクション |
|---|---|---|
|
|
|
| Item type | 学術雑誌論文 / Journal Article(1) | |||||
|---|---|---|---|---|---|---|
| 公開日 | 2017-10-03 | |||||
| タイトル | ||||||
| タイトル | Graphite Furnace Atomic Absorption Spectrometric Determination of Bismuth(III) after Coprecipitation with Hafnium Hydroxide | |||||
| 言語 | ||||||
| 言語 | eng | |||||
| 資源タイプ | ||||||
| 資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
| 資源タイプ | journal article | |||||
| 著者 |
Ueda, Joichi
× Ueda, Joichi× Takagi, Midori |
|||||
| 書誌情報 |
Bulletin of the Chemical Society of Japan 巻 63, 号 2, p. 544-547, 発行日 1990-01-01 |
|||||
| ISSN | ||||||
| 収録物識別子タイプ | ISSN | |||||
| 収録物識別子 | 0009-2673 | |||||
| NCID | ||||||
| 収録物識別子タイプ | NCID | |||||
| 収録物識別子 | AA00580132 | |||||
| DOI | ||||||
| 関連タイプ | isIdenticalTo | |||||
| 識別子タイプ | DOI | |||||
| 関連識別子 | 10.1246/bcsj.63.544 | |||||
| 出版者 | ||||||
| 出版者 | The Chemical Society of Japan = 日本化学会 | |||||
| 抄録 | ||||||
| 内容記述タイプ | Abstract | |||||
| 内容記述 | A method for the coprecipitation of bismuth(III) with hafnium hydroxide followed by a graphite-furnace atomic absorption spectrometric determination is described. Hafnium hydroxide coprecipitates quantitatively 0.05–3 μg of bismuth(III) from 50–400 cm3 of sample solution at pH 5.8–11.2. The presence of 2.5–50 mg of hafnium in 25 cm3 does not affect the atomic absorbance of bismuth(III). The calibration curve is linear for 0.05–3 μg of bismuth(III) in 25 cm3 and passes through the origin. Interference from the presence of large amounts of silver can be eliminated by removing silver as the ammine complex produced by the coprecipitation procedure at above pH 11 with aqueous ammonia. Several other matrix ions investigated do not interfere with the determination in at least a 500-fold mass ratio to bismuth(III). This method was successfully applied to the determination of trace bismuth(III) in commercial metal silver (purity: 99.99%), and is also applicable to the determination of trace amounts of bismuth(III) in water sample. | |||||
| 権利 | ||||||
| 権利情報 | Copyright © The Chemical Society of Japan 日本化学会 | |||||
| 著者版フラグ | ||||||
| 出版タイプ | VoR | |||||
| 出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||
| 関連URI | ||||||
| 識別子タイプ | URI | |||||
| 関連識別子 | https://www.jstage.jst.go.jp/browse/bcsj | |||||
| 関連URI | ||||||
| 識別子タイプ | URI | |||||
| 関連識別子 | http://www.chemistry.or.jp/ | |||||