Accuracy of ventricular volume and ejection fraction measured by gated myocardial SPECT: Comparison of 4 software programs

http://hdl.handle.net/2297/2797
http://hdl.handle.net/2297/2797
29ebf0f5-7018-4709-98e3-bace013e7e5d
名前 / ファイル ライセンス アクション
ME-PR-NAKAJIMA-K-1571.pdf ME-PR-NAKAJIMA-K-1571.pdf (178.5 kB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-10-03
タイトル
タイトル Accuracy of ventricular volume and ejection fraction measured by gated myocardial SPECT: Comparison of 4 software programs
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Nakajima, Kenichi

× Nakajima, Kenichi

WEKO 320
e-Rad 00167545
金沢大学研究者情報 00167545
研究者番号 00167545

Nakajima, Kenichi

Search repository
Higuchi, Takahiro

× Higuchi, Takahiro

WEKO 20095

Higuchi, Takahiro

Search repository
Taki, Junichi

× Taki, Junichi

WEKO 264
e-Rad 00167545
金沢大学研究者情報 10251927
研究者番号 10251927

Taki, Junichi

Search repository
Kawano, Masaya

× Kawano, Masaya

WEKO 20096

Kawano, Masaya

Search repository
Tonami, Norihisa

× Tonami, Norihisa

WEKO 20066
e-Rad 60019940
研究者番号 60019940

Tonami, Norihisa

Search repository
提供者所属
内容記述タイプ Other
内容記述 金沢大学大学院医学系研究科
書誌情報 Journal of Nuclear Medicine

巻 42, 号 10, p. 1571-1578, 発行日 2001-10-01
ISSN
収録物識別子タイプ ISSN
収録物識別子 0161-5505
NCID
収録物識別子タイプ NCID
収録物識別子 AA00703684
出版者
出版者 THE SOCIETY OF NUCLEAR MEDICINE INC
抄録
内容記述タイプ Abstract
内容記述 Gated myocardial perfusion SPECT has been used to calculate ejection fraction (EF) and end-diastolic volume (EDV) and has correlated well with conventional methods. However, the comparative accuracy of and correlations across various types of gated SPECT software are not well understood. Methods: Mathematic phantoms of cylindric-hemispheric hybrid models, ranging in volume from 34 to 266 mL, were generated. The clinical cases consisted of 30 patients who participated in a radionuclide angiography and gated blood-pool (GBP) study in addition to undergoing 99mTc-sestamibi gated SPECT. Four kinds of software, Quantitative Gated SPECT (QGS), the Emory Cardiac Toolbox (ECT), 4D-MSPECT, and Perfusion and Functional Analysis for Gated SPECT (pFAST) were used to compute EF and EDV, and the results were analyzed by multiple comparisons tests. Patients were classified into 4 groups (i.e., no defect, small defect, large defect, and small heart) so that factors affecting variation could be analyzed. Results: In mathematic models ≥74 mL, volume error was within ±15%, whereas for a small volume (34 mL), QGS and 4D-MSPECT underestimated the volume and pFAST overestimated it. The respective intra- and interobserver reproducibility of the results was good for QGS (r=0.99 and 1.00), ECT (r=0.98 and 0.98), and 4D-MSPECT (r=0.98 and 0.98) and fair for pFAST (r=0.88 and 0.85). The correlation coefficient for EF between gated SPECT and the GBP study was 0.82, 0.78, 0.69, and 0.84 for QGS, ECT, 4D-MSPECT, and pFAST, respectively. The correlation coefficient for EDV between gated SPECT and the GBP study was 0.88, 0.89, 0.85, and 0.90, respectively. Although good correlation was observed among the 4 software packages, QGS, ECT, and 4D-MSPECT overestimated EF in patients with small hearts, and pFAST overestimated the true volume in patients with large perfusion defects. Correlation coefficients among the 4 kinds of software were 0.80-0.95 for EF and 0.89-0.98 for EDV. Conclusion: All 4 software programs showed good correlation between EF or EDV and the GBP study. Good correlation was observed also between each pair of quantification methods. However, because each method has unique characteristics that depend on its specific algorithm and thus behaves differently in the various patient subgroups, the methods should not be used interchangeably.
著者版フラグ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
戻る
0
views
See details
Views

Versions

Ver.1 2023-07-27 09:48:14.402177
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3