北陸地域の訪問線量測定による基準点吸収線量測定とQA/QCアンケート調査(第1報) : 過去データとの比較による経年的推移について
http://hdl.handle.net/2297/25349
http://hdl.handle.net/2297/25349a594c1be-1947-4388-9879-fd241a1f0390
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-10-03 | |||||
タイトル | ||||||
タイトル | 北陸地域の訪問線量測定による基準点吸収線量測定とQA/QCアンケート調査(第1報) : 過去データとの比較による経年的推移について | |||||
タイトル | ||||||
タイトル | Result of the onsite dosimetry and the questioner about quality assurance/quality control for radiotherapy in Hokuriku area. - A comparison with the past results – | |||||
言語 | en | |||||
言語 | ||||||
言語 | jpn | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
著者 |
武村, 哲浩
× 武村, 哲浩× 塚本, 直孝× 山本, 瑛一× 上田, 伸一× 倉田, 雄一× 嘉戸, 祥介× 西島, 昭彦× 松倉, 昭芳× 越田, 吉郎× 天野, 良平 |
|||||
書誌情報 |
日本放射線技術學會雜誌 巻 66, 号 9, p. 1213-1220, 発行日 2010-01-01 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0369-4305 | |||||
NCID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AN00197784 | |||||
出版者 | ||||||
出版者 | 日本放射線技術学会 | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Purpose: To analyze temporal changes in human resources in the radiotherapy section, quality assurance/quality control (QA/QC) and dose difference for radiotherapy in the Hokuriku area based on the results of past investigations and our investigation. Method: We visited radiotherapy sections of 17 hospitals in the Hokuriku area (5 in Toyama, 9 in Ishikawa and 3 in Fukui) to measure the dose at the reference point of a linear accelerator (LINAC), as we asked questions to a radiotherapist about human resources, QA/QC of LINAC, etc. We compared our results with past reports (1992 to 2007) on the dose difference, human resources and frequency of dose monitor system calibration. Results: The number of physicians has not changed since 1999, but the number of radiotherapists was significantly increased. Weekly dose monitor system calibration has been achieved in 80% of the institutions in our survey. This percentage was significantly higher than in the past surveys. The dose difference distribution from our onsite dosimetry did not significantly differ from that from the onsite dosimetry in 2007. 91% of the institutions have accomplished within 2% of the dose difference. Conclusion: We found that the number of physicians has not increased since 1999, but the number of radiotherapists has increased. We conclude that the increment of radiotherapists led to 80% achievement of the weekly dose monitor system calibration. Almost all institutions in Hokuriku area have properly performed QA of the dose monitor system. | |||||
権利 | ||||||
権利情報 | Copyright (c) 2010 社団法人 日本放射線技術学会 | |||||
著者版フラグ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | http://www.jsrt.or.jp/mokuji.pdf | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | http://www.jstage.jst.go.jp/article/jjrt/66/9/66_1213/_article/-char/ja/ |