@article{oai:kanazawa-u.repo.nii.ac.jp:00013735, author = {五味, 勉 and 春日, 敏夫 and 角谷, 眞澄 and 宮地, 利明 and 越田, 吉郎 and 安藤, 隆 and 上田, 仁 and 佐々木, 祥之}, issue = {3}, journal = {医用画像情報学会雑誌 = Medical Imaging and Information Sciences}, month = {Jan}, note = {We analysed of the images generated with multi-slice CT in the longitudinal direction (Z) and assessed the noise equivalent quanta (NEQ)evaluation. Evaluation of image quality in NEQ, was possible to accuracy of reappearance in low at high frequency characteristics. In addition, NEQ information was confirmed of usefulness in compared with Winer spectra and MTF method. Examination results suggest that the quantitative analysis in characteristics of image noise and image resolution at multi-slice CT images can provide an optimal parameter for improving quality of images in clinical data.}, pages = {136--140}, title = {NEQを用いたマルチスライスCTの体軸方向における画質評価}, volume = {19}, year = {2002} }