@article{oai:kanazawa-u.repo.nii.ac.jp:00014226, author = {Hara, Takanori and Urikura, Atsushi and Ichikawa, Katsuhiro and Hoshino, Takashi and Nishimaru, Eiji and Niwa, Shinji}, issue = {4}, journal = {Physica Medica}, month = {Apr}, note = {Purpose: To analyse the temporal resolution (TR) of modern computed tomography (CT) scanners using the impulse method, and assess the actual maximum TR at respective helical acquisition modes. Methods: To assess the actual TR of helical acquisition modes of a 128-slice dual source CT (DSCT) scanner and a 320-row area detector CT (ADCT) scanner, we assessed the TRs of various acquisition combinations of a pitch factor (P) and gantry rotation time (R). Results: The TR of the helical acquisition modes for the 128-slice DSCT scanner continuously improved with a shorter gantry rotation time and greater pitch factor. However, for the 320-row ADCT scanner, the TR with a pitch factor of <1.0 was almost equal to the gantry rotation time, whereas with pitch factor of >1.0, it was approximately one half of the gantry rotation time. The maximum TR values of single- and dual-source helical acquisition modes for the 128-slice DSCT scanner were 0.138 (R/. P = 0.285/1.5) and 0.074. s (R/. P = 0.285/3.2), and the maximum TR values of the 64. ×. 0.5- and 160. ×. 0.5-mm detector configurations of the helical acquisition modes for the 320-row ADCT scanner were 0.120 (R/. P = 0.275/1.375) and 0.195. s (R/. P = 0.3/0.6), respectively. Conclusion: Because the TR of a CT scanner is not accurately depicted in the specifications of the individual scanner, appropriate acquisition conditions should be determined based on the actual TR measurement. © 2016 Associazione Italiana di Fisica Medica.}, pages = {625--630}, title = {Temporal resolution measurement of 128-slice dual source and 320-row area detector computed tomography scanners in helical acquisition mode using the impulse method}, volume = {32}, year = {2016} }