@article{oai:kanazawa-u.repo.nii.ac.jp:00014766, author = {石崎, 純子 and 横川, 弘一 and 中島, 恵美 and 古田, 壽一 and 越野, 好文 and 関, 秀俊 and 谷口, 昂 and 長谷川, 光広 and 山下, 純宏 and 市村, 藤雄}, issue = {4}, journal = {病院薬学}, month = {Aug}, note = {A fully automated enzyme immunoassay system, Omni (BIO-TEK Co.) (ELISA system), using Markit-M Excegran and Markit-M Haloperidol (Dainippon Pharmaceutical Co.) was evaluated to determine the concentrations of zonisamide (ZNS) and haloperidol (HP) in human plasma. At first, the within-run (ZNS : n=8, HP : n=10) and the between-run (4 days) precisions of ZNS and HP were examined. The coefficient of the variations in both the within-run and the between-run previsions of HP were below 5.0 and 7.00/0, respectively. In contrast, the some values of ZNS were below 6.5 and 8.40/0, respectively. A good correlation was observed between this ELISA system and HPLC (r=0.917), which was used to determine the plasma ZNS concentration. In the case of HP, a good correlation was seen between this ELISA system and the manual ELISA method (r=0.978). This method was thus used to monitor the serum levels in patients receiving ZNS or HP therapy. The relationship was observed between the daily dose (mg/kg) and the plasma concentration of the trough level for ZNS and HP, respectively (r=0.750 p<0.001, r= 0.669 p<0.001). However, the ratio of the plasma concentration of ZNS and the daily dose increased significantly with aging., 金沢大学大学院自然科学研究科分子作用学, 金沢大学医学部附属病院薬剤部, 金沢大学大学院医学系研究科}, pages = {358--363}, title = {全自動ELISAシステムによるゾニサミドおよびハロペリドールの血漿中濃度測定}, volume = {23}, year = {1997} }