Reducing the small-heart effect in pediatric gated myocardial perfusion single-photon emission computed tomography

http://hdl.handle.net/2297/45932
http://hdl.handle.net/2297/45932
6befe847-8e76-4743-afd0-51220ddbfcfc
名前 / ファイル ライセンス アクション
HO-PR-YONEYAMA-H-1378.pdf HO-PR-YONEYAMA-H-1378.pdf (870.2 kB)
 Download is available from 9999/12/31.
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-11-09
タイトル
タイトル Reducing the small-heart effect in pediatric gated myocardial perfusion single-photon emission computed tomography
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Yoneyama, Hiroto

× Yoneyama, Hiroto

WEKO 66281

Yoneyama, Hiroto

Search repository
Nakajima, Kenichi

× Nakajima, Kenichi

WEKO 320
e-Rad 00167545
金沢大学研究者情報 00167545
研究者番号 00167545

Nakajima, Kenichi

Search repository
Okuda, Koichi

× Okuda, Koichi

WEKO 66282

Okuda, Koichi

Search repository
Matsuo, Shinro

× Matsuo, Shinro

WEKO 599
金沢大学研究者情報 30359773
研究者番号 30359773

Matsuo, Shinro

Search repository
Onoguchi, Masahisa

× Onoguchi, Masahisa

WEKO 352
金沢大学研究者情報 30283120
研究者番号 30283120

Onoguchi, Masahisa

Search repository
Kinuya, Seigo

× Kinuya, Seigo

WEKO 115
e-Rad 20281024
金沢大学研究者情報 20281024
研究者番号 20281024

Kinuya, Seigo

Search repository
Edenbrandt, Lars

× Edenbrandt, Lars

WEKO 66283

Edenbrandt, Lars

Search repository
書誌情報 Journal of Nuclear Cardiology

巻 24, 号 4, p. 1378-1388, 発行日 2017-08-01
ISSN
収録物識別子タイプ ISSN
収録物識別子 1071-3581
NCID
収録物識別子タイプ NCID
収録物識別子 AA11016223
DOI
関連タイプ isVersionOf
識別子タイプ DOI
関連識別子 10.1007/s12350-016-0518-z
出版者
出版者 Springer New York LLC
抄録
内容記述タイプ Abstract
内容記述 Background: We compared two reconstruction algorisms and two cardiac functional evaluation software programs in terms of their accuracy for estimating ejection fraction (EF) of small hearts (SH). Methods: The study group consisted of 66 pediatric patients. Data were reconstructed using a filtered back projection (FBP) method without the resolution correction (RC) and an iterative method based on an ordered subset expectation maximization (OSEM) algorithm with the RC. EF was evaluated using two software programs of quantitative gated single-photon emission computed tomography (SPECT) (QGS) and cardioREPO. We compared the EF of gated myocardial perfusion SPECT to echocardiographic measurement (Echo). Results: Forty-eight of 66 patients had an end-systolic volume < 20 mL which was used as the criterion for being included in the SH group, and the SH effect led to an overestimation of EF. While significant differences were observed between Echo (66.9 ± 5.0%) and QGS-FBP without RC (76.9 ± 8.4%, P < .0001), QGS-OSEM with RC (76.6 ± 8.6%, P < .0001), and cardioREPO-FBP without RC (72.1 ± 10.0%, P = .0011), no significant difference was observed between Echo and cardioREPO-OSEM with RC (67.4 ± 6.1%) in SH group. Conclusions: In pediatric gated myocardial perfusion SPECT, the SH effect can be significantly reduced when an OSEM algorithm is used with RC in combination with the specific cardioREPO algorithm. © 2016 American Society of Nuclear Cardiology
内容記述
内容記述タイプ Other
内容記述 Embargo Period 12 months
著者版フラグ
出版タイプ AM
出版タイプResource http://purl.org/coar/version/c_ab4af688f83e57aa
戻る
0
views
See details
Views

Versions

Ver.1 2023-07-27 17:53:22.882214
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3