@article{oai:kanazawa-u.repo.nii.ac.jp:00044478, author = {Misaki, Takuro and Ohtake, Hiroshi and Matsumoto, Isao and Matsunaga, Yasuhiro and Watanabe, Yoh and 三崎, 拓郎 and 大竹, 裕志 and 松本, 勲 and 松永, 康弘 and 渡辺, 洋宇}, issue = {9}, journal = {胸部外科 = 日本心臓血管外科学会雑誌, The Japanese journal of thoracic surgery}, month = {Aug}, note = {To enhance the utility of epicardial mapping in the intraoperative determination of the location of the arrhythmic source, we developed a new balloon electrode covered with Presnet-tubular dressing material including 46 snap electrode recording sites. The bipolar recording sites, 1.0 mm in diameter and spaced at intervals of 1.5 mm, are constructed of gold and directly attached steel wire. This new balloon electrode was effective intraoperative determination of the lesion in 2 patients one with Wolff-Parkinson-White syndrome and the other with ventricular tachycardia., 金沢大学医薬保健研究域医学系}, pages = {794--796}, title = {スナップ電極を応用した心内膜バルーン電極の開発}, volume = {45}, year = {1992} }