@techreport{oai:kanazawa-u.repo.nii.ac.jp:00052208, month = {Apr}, note = {In this study, we challenged to develop a method for verifying the accuracy of the accumulated dose distribution calculated with the deformable image registration (DIR) in radiotherapy. We use a gel dosimeter that can be measured three-dimensionally and can be deformed. We irradiated the gel dosimeter before and after the deformation, also scanned it by CT scanner and build each treatment plan, then the accu,ulated dose distribution by DIR was calculated. The accumulated dose distribution was compared with the dose distribution read from the gel dosimeter. Since the sensitivity of the gel dosimeter decreases when gel dosimeter is exposed to oxygen, an oxygen blocking method for the verification method could be developed. And it was found that the gelatin of the gel dosimeter collapsed due to the deformation of the gel dosimeter and the sensitivity of the collapsed part decreased., 本研究では, 放射線治療におけるDeformable image registration(DIR)を用いた合算線量分布の精度検証方法開発のために,3次元的に測定でき変形可能なゼリー状のゲル線量計を用いることを検討した.ゲル線量計を変形前と変形後に照射し, それぞれのCT撮影とそれぞれの治療計画を立てDIRによる合算線量分布をえてゲル線量計から読み出した線量分布を比較する評価方法を検討した.ゲル線量計は酸素に触れると感度が低下するため,検証方法に用いる酸素遮断方法が開発できたが,ゲル線量計の変形によってゲル線量計のゼラチンが崩れ,崩れた部分の感度が低下することがわかった., 研究課題/領域番号:19K12818, 研究期間(年度):2019-04-01 – 2022-03-31, 出典:研究課題「放射線治療における変形画像照合技術を用いた線量合算精度の評価手法確立」課題番号19K12818 (KAKEN:科学研究費助成事業データベース(国立情報学研究所)) (https://kaken.nii.ac.jp/ja/report/KAKENHI-PROJECT-19K12818/19K12818seika/)を加工して作成, 金沢大学医薬保健研究域保健学系}, title = {放射線治療における変形画像照合技術を用いた線量合算精度の評価手法確立}, year = {2022} }