Cable externalization at the proximal portion of the superior vena cava coil in Riata implantable cardioverter defibrillator leads
http://hdl.handle.net/2297/46782
http://hdl.handle.net/2297/46782fb42061c-b667-412c-9208-c46f1cebb4e0
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-12-05 | |||||
タイトル | ||||||
タイトル | Cable externalization at the proximal portion of the superior vena cava coil in Riata implantable cardioverter defibrillator leads | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
著者 |
Katsuki, Tomonori
× Katsuki, Tomonori× Furusho, Hiroshi× Kusayama, Takashi× Takashima, Shinichiro× Kato, Takeshi× Murai, Hisayoshi× Usui, Soichiro× Kaneko, Shuichi× Takamura, Masayuki |
|||||
書誌情報 |
Journal of Cardiology Cases 巻 15, 号 1, p. 3-5, 発行日 2017-01-01 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 1878-5409 | |||||
DOI | ||||||
関連タイプ | isVersionOf | |||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1016/j.jccase.2016.08.009 | |||||
出版者 | ||||||
出版者 | 日本心臓病学会 = Japanese College of Cardiology / Elsevier | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Many Riata (St. Jude Medical, St. Paul, MN, USA) implantable cardioverter defibrillator (ICD) leads have reportedly developed cable externalization. The most likely cause of cable externalization is insulation abrasion, which often occurs at the can or between the right ventricular coil and superior vena cava (SVC) coil. We report a rare case of an adult male whose ICD lead cable was externalized at the proximal portion of the SVC coil. This lead became fixed to the wall at the subclavian vein and SVC and became bent between these adhesions. Furthermore, the motion of this lead was affected by pulsation of the aortic arch. The ICD lead might develop inside-out abrasion due to mechanical stress evoked by pulsation of the aortic arch at this site.<. Learning objective: Cable externalization of the implantable cardioverter defibrillator lead at the proximal portion of the superior vena cava (SVC) coil has rarely been reported. Externalization might be the result of deformation of the left brachiocephalic vein and the anatomical relationship with the aortic arch. The anatomical pathway of the lead should be carefully considered during the procedure, especially when a dual-coil lead is selected. Moreover, possible cable externalization at both the proximal and distal portions of the SVC coil should be kept in mind during follow-up>. © 2016 Japanese College of Cardiology. | |||||
内容記述 | ||||||
内容記述タイプ | Other | |||||
内容記述 | Embargo Period 12 months | |||||
権利 | ||||||
権利情報 | Copyright © Japanese College of Cardiology / Elsevier (CC-BY NC ND): This is the author's version of a work accepted for publication by Japanese College of Cardiology. Changes resulting from the publishing process, including peer review, editing, corrections, structural formatting and other quality control mechanisms, may not be reflected in this document. Changes may have been made to this work since it was submitted for publication. The definitive version has been published in Journal of Cardiology Cases,15, 1, 1 January 2017, 10.1016/j.jccase.2016.08.009. | |||||
著者版フラグ | ||||||
出版タイプ | AM | |||||
出版タイプResource | http://purl.org/coar/version/c_ab4af688f83e57aa | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | http://www.elsevier.com/locate/issn/18785409 | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | http://www.jcc.gr.jp/ |