WEKO3
インデックスリンク
アイテム
肝動脈でのroadmap機能における横隔膜位置を用いた呼吸性移動の補正手法の信頼性評価
http://hdl.handle.net/2297/27690
http://hdl.handle.net/2297/276907008aefa-be33-4fab-92f1-bf7f1f7419e4
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-10-03 | |||||
タイトル | ||||||
タイトル | 肝動脈でのroadmap機能における横隔膜位置を用いた呼吸性移動の補正手法の信頼性評価 | |||||
タイトル | ||||||
タイトル | Reliability of the roadmap technique with the respiratory motion compensation using hemi-diaphragm position for hepatic arteries | |||||
言語 | en | |||||
言語 | ||||||
言語 | jpn | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
著者 |
武村, 哲浩
× 武村, 哲浩× 三井, 渉× 茶畠, 光浩× 飯田, 泰治× 奥村, 悠祐× 鈴木, 正行× 原内, 一 |
|||||
書誌情報 |
医用画像情報学会雑誌 = Japan Society of Imaging and Information Sciences in Medicine 巻 27, 号 1, p. 1-5, 発行日 2010-01-01 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0910-1543 | |||||
NCID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AN10156808 | |||||
出版者 | ||||||
出版者 | 医用画象情報学会 | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Miss-placement of arteries while using the roadmap technique in endovascular interventions for abdominal arteries occurs because of respiration. It has been reported that abdominal arteries move mainly in a superior-inferior direction and that diaphragm position helps to compensate for the miss-placement. We investigated the reliability of a roadmap technique with respiratory motion compensation that used diaphragm positions to cancel out any miss-placement of the hepatic arteries. From three patients, we obtained five pairs of fluoroscopic sequences and digital subtraction angiography (DSA) images. The coordinates of the right hemi-diaphragm and a micro guide-wire tip were noted for every frame of the fluoroscopic sequence along with the coordinates of the right hemi-diaphragm on the DSA image. The micro guide-wire tip coordinates were corrected using the difference of the Y coordinate values from both hemi-diaphragms. The tip positions that were successfully placed on arteries in the DSA image were counted and the success rate calculated. Mean success rates for two cases were over 90%, though that for the other was under 10%. The results indicate that reliability of the roadmap technique with respiratory motion compensation depends on the particular case. However, this method could stabilize the micro guide-wire tip position. | |||||
権利 | ||||||
権利情報 | Copyright (c) 2010 医用画像情報学会 | |||||
著者版フラグ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | http://www.mii-sci.jp/ | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | http://ci.nii.ac.jp/naid/130000262784 | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | http://www.jstage.jst.go.jp/article/mii/27/1/27_1/_article |