WEKO3
インデックスリンク
アイテム
高速スピンエコー法を用いたT1・T2同時計測MRI
http://hdl.handle.net/2297/36646
http://hdl.handle.net/2297/36646a590892e-de39-4a14-8e06-ceac3169ad12
名前 / ファイル | ライセンス | アクション |
---|---|---|
ME-PR-MITATI-T-111.pdf (987.4 kB)
|
|
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-10-03 | |||||
タイトル | ||||||
タイトル | 高速スピンエコー法を用いたT1・T2同時計測MRI | |||||
タイトル | ||||||
タイトル | T1 and T2 Hybrid Measurements using Fast Spin-echo Sequences | |||||
言語 | en | |||||
言語 | ||||||
言語 | jpn | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
著者 |
金澤, 裕樹
× 金澤, 裕樹× 宮地, 利明× 八木, 絢子× 吉田, 彩× 佐藤, 修 |
|||||
著者別表示 |
Kanazawa, Yuki
× Kanazawa, Yuki× Miyati, Toshiaki× Yagi, Ayako× Yoshida, Aya× Sato, Osamu |
|||||
書誌情報 |
医用画像情報学会雑誌 = Medical Imaging and Information Sciences 巻 27, 号 4, p. 111-115, 発行日 2010-01-01 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0910-1543 | |||||
NCID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AN10156808 | |||||
DOI | ||||||
関連タイプ | isIdenticalTo | |||||
識別子タイプ | DOI | |||||
関連識別子 | 10.11318/mii.27.111 | |||||
出版者 | ||||||
出版者 | 医用画像情報学会 = MII: Japan Society of Medical Imaging and Information Sciences | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | The purpose of our study is to develop and to assess a method to calculate T1 and T2 relaxation times using fast spin-echo (FSE) sequences with two echoes per excitation. On a 1.5-T MRI, dual-echo FSE sequences with different repetition times (TR:TR2=2TR1, i.e., TR1=400ms and TR2=800ms) were performed in MnCl2 phantom. Then, we assessed effects of the echo train length (ETL) , effective echo time (TE) , and T2 correction for T1 measurement, when changed each imaging parameter. Besides, we evaluated with comparison between the dual-echo FSE method and mixed sequence. With optimizing FSE imaging parameters, i.e., effective TE, TR, and low ETL, the measurement values of T1 and T2 revealed significantly higher correlation between the dual FSE method and mixed sequence. Moreover, the T2 correction to calculate T1 value could be omitted by shortening first effective TE. The dual-echo FSE method makes it possible to measure T1 and T2 fast and easily and to obtain more detailed information of organizations in human. | |||||
権利 | ||||||
権利情報 | Copyright © 医用画像情報学会 MII: Japan Society of Medical Imaging and Information Sciences | |||||
著者版フラグ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://www.jstage.jst.go.jp/browse/mii/-char/ja/ | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | http://www.mii-sci.jp/ |