Where should we measure the entrance air kerma rate during acceptance testing of the automatic dose control of a fluoroscopic system?

http://hdl.handle.net/2297/34134
http://hdl.handle.net/2297/34134
a4333bb0-171e-4cb0-af72-8d43fbb59192
名前 / ファイル ライセンス アクション
ME-PR-FUKUDA-A-313.pdf ME-PR-FUKUDA-A-313.pdf (152.0 kB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-10-03
タイトル
タイトル Where should we measure the entrance air kerma rate during acceptance testing of the automatic dose control of a fluoroscopic system?
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Fukuda, Atsushi

× Fukuda, Atsushi

WEKO 23156

Fukuda, Atsushi

Search repository
Miyati, Tosiaki

× Miyati, Tosiaki

WEKO 120
e-Rad 80324086
金沢大学研究者情報 80324086
研究者番号 80324086

Miyati, Tosiaki

Search repository
Matsubara, Kosuke

× Matsubara, Kosuke

WEKO 521
金沢大学研究者情報 30507372
研究者番号 30507372

Matsubara, Kosuke

Search repository
書誌情報 Radiological Physics and Technology

巻 6, 号 2, p. 313-316, 発行日 2013-07-01
ISSN
収録物識別子タイプ ISSN
収録物識別子 1865-0333
NCID
収録物識別子タイプ NCID
収録物識別子 AA12236881
DOI
関連タイプ isVersionOf
識別子タイプ DOI
関連識別子 10.1007/s12194-013-0202-9
出版者
出版者 Japanese Society of Radiological Technology 日本放射線技術学会 / Springer Verlag (Germany)
抄録
内容記述タイプ Abstract
内容記述 In Japan, the entrance air kerma rate (EAKR) to a patient cannot exceed 50 mGy/min in conventional fluoroscopy. However, it is unclear where the EAKR should be measured. We obtained the tube potential and tube current as a function of polymethylmethacrylate (PMMA) thickness, and the EAKR at the interventional reference point (IRP) was measured from the trajectory. The EAKR at the point established by the U.S. Food and Drug Administration (FDA) was calculated from EAKR at the IRP. The EAKR at the IRP exceeded the limit at a PMMA thickness of 22-28 cm. However, the EAKR did not exceed the limit at the FDA point. If the EAKR to a patient is being verified to meet the recent Japanese ruling, the EAKR should be measured at the FDA point, and if the EAKR is being evaluated for determination of the skin dose, it should be monitored at the IRP. © 2013 Japanese Society of Radiological Technology and Japan Society of Medical Physics.
著者版フラグ
出版タイプ AM
出版タイプResource http://purl.org/coar/version/c_ab4af688f83e57aa
戻る
0
views
See details
Views