Measurement of table feed speed in modern CT

http://hdl.handle.net/2297/37480
http://hdl.handle.net/2297/37480
b5255d51-1421-4c1f-82e5-e74a61759653
名前 / ファイル ライセンス アクション
ME-PR-FUKUDA-A-275.pdf ME-PR-FUKUDA-A-275.pdf (669.2 kB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-10-03
タイトル
タイトル Measurement of table feed speed in modern CT
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Fukuda, Atsushi

× Fukuda, Atsushi

WEKO 23397

Fukuda, Atsushi

Search repository
Lin, Pei-Jan P.

× Lin, Pei-Jan P.

WEKO 23398

Lin, Pei-Jan P.

Search repository
Matsubara, Kosuke

× Matsubara, Kosuke

WEKO 521
金沢大学研究者情報 30507372
研究者番号 30507372

Matsubara, Kosuke

Search repository
Miyati, Tosiaki

× Miyati, Tosiaki

WEKO 120
e-Rad 80324086
金沢大学研究者情報 80324086
研究者番号 80324086

Miyati, Tosiaki

Search repository
書誌情報 Journal of Applied Clinical Medical Physics

巻 15, 号 3, p. 275-281, 発行日 2014-01-01
ISSN
収録物識別子タイプ ISSN
収録物識別子 1526-9914
NCID
収録物識別子タイプ NCID
収録物識別子 AA12034060
DOI
関連タイプ isIdenticalTo
識別子タイプ DOI
関連識別子 10.1120/jacmp.v15i3.4703
出版者
出版者 American Association of Physicists in Medicine
抄録
内容記述タイプ Abstract
内容記述 The purpose of this study was to develop and evaluate a noninvasive method to assess table feed speed (mm/s) in modern commercial computed tomography (CT) systems. The table feed (mm/rotation) was measured at selected nominal table feed speeds, given as low (26.67 mm/s), intermediate (48.00 mm/s), and high (64.00mm/s), by utilizing a computed radiography (CR) cassette installed with a photostimulable phosphor plate. The cassette was placed on the examination table to travel through the isocenter longitudinally, with a total scan length of over 430mm. The distance travelled was employed to determine the total table feed length. To calculate the table feed speed, gantry rotation time was measured concurrently at a preselected nominal rotation time of 750 ms. Upon completion of data acquisition, the table feed and gantry rotation time were analyzed and used to calculate the actual table feed speed (mm/s). Under the low table feed speed setting, the table feed speed was found to be 26.67 mm/s. Similarly, under the intermediate and high table feed speed settings, the table feed speed was found to be 48.10 and 64.07mm/s, respectively. Measurements of the table feed speed can be accomplished with a CR system and solid-state detector, and the table feed speed results were in excellent agreement with the nominal preset values.
著者版フラグ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
関連URI
識別子タイプ URI
関連識別子 http://www.jacmp.org/index.php/jacmp/article/view/4703
戻る
0
views
See details
Views

Versions

Ver.1 2023-07-28 00:57:16.838102
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3