静磁場強度の異なる3機種を用いた脳定位放射線治療計画のためのMRI画像の歪み比較
http://hdl.handle.net/2297/35102
http://hdl.handle.net/2297/351029a78d163-d226-40a9-8e70-6472264c161b
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-10-03 | |||||
タイトル | ||||||
タイトル | 静磁場強度の異なる3機種を用いた脳定位放射線治療計画のためのMRI画像の歪み比較 | |||||
タイトル | ||||||
タイトル | Comparison of Image Distortion between Three Magnetic Resonance Imaging Systems of Different Magnetic Field Strengths for Use in Stereotactic Irradiation of Brain | |||||
言語 | en | |||||
言語 | ||||||
言語 | jpn | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
著者 |
武村, 哲浩
× 武村, 哲浩× 笹本, 耕平× 仲村, 香織× 黒田, 辰憲× 小路, 佐織× 松浦, 幸広× 松下, 達彦 |
|||||
書誌情報 |
日本放射線技術学会雑誌 = Japanese journal of radiological technology 巻 69, 号 6, p. 641-647, 発行日 2013-01-01 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0369-4305 | |||||
NCID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AN00197784 | |||||
DOI | ||||||
関連タイプ | isIdenticalTo | |||||
識別子タイプ | DOI | |||||
関連識別子 | 10.6009/jjrt.2013_JSRT_69.6.641 | |||||
出版者 | ||||||
出版者 | 日本放射線技術学会 = Japanese Society of Radiological Technology | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | In this study, we evaluated the image distortion of three magnetic resonance imaging (MRI) systems with magnetic field strengths of 0.4 T, 1.5 T and 3 T, during stereotactic irradiation of the brain. A quality assurance phantom for MRI image distortion in radiosurgery was used for these measurements of image distortion. Images were obtained from a 0.4-T MRI (APERTO Eterna, HITACHI), a 1.5-T MRI (Signa HDxt, GE Healthcare) and a 3-T MRI (Signa HDx 3.0 T, GE Healthcare) system. Imaging sequences for the 0.4-T and 3-T MRI were based on the 1.5-T MRI sequence used for stereotactic irradiation in the clinical setting. The same phantom was scanned using a computed tomography (CT) system (Aquilion L/B, Toshiba) as the standard. The results showed mean errors in the Z direction to be the least satisfactory of all the directions in all results. The mean error in the Z direction for 1.5-T MRI at -110 mm in the axial plane showed the largest error of 4.0 mm. The maximum errors for the 0.4-T and 3-T MRI were 1.7 mm and 2.8 mm, respectively. The errors in the plane were not uniform and did not show linearity, suggesting that simple distortion correction using outside markers is unlikely to be effective. The 0.4-T MRI showed the lowest image distortion of the three. However, other items, such as image noise, contrast and study duration need to be evaluated in MRI systems when applying frameless stereotactic irradiation. | |||||
権利 | ||||||
権利情報 | Copyright © 2013 公益社団法人 日本放射線技術学会 / 許可を得て登録 | |||||
著者版フラグ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | http://www.jsrt.or.jp/ | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://www.jstage.jst.go.jp/article/jjrt/69/6/69_2013_JSRT_69.6.641/_article/-char/ja/ | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://japanlinkcenter.org/DN/JST.JSTAGE/jjrt/2013_JSRT_69.6.641 |