A novel removable shield attached to C-arm units against scattered X-rays from a patient's side

http://hdl.handle.net/2297/39048
http://hdl.handle.net/2297/39048
fd787266-6fd7-4fb2-a7b9-0d418ee1ae72
名前 / ファイル ライセンス アクション
ME-PR-MORI-H-1794.pdf ME-PR-MORI-H-1794.pdf (400.4 kB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-10-03
タイトル
タイトル A novel removable shield attached to C-arm units against scattered X-rays from a patient's side
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Mori, Hiroshige

× Mori, Hiroshige

WEKO 24136

Mori, Hiroshige

Search repository
Koshida, Kichiro

× Koshida, Kichiro

WEKO 20089
e-Rad 90020023
研究者番号 90020023

Koshida, Kichiro

Search repository
Ishigamori, Osamu

× Ishigamori, Osamu

WEKO 24137

Ishigamori, Osamu

Search repository
Matsubara, Kosuke

× Matsubara, Kosuke

WEKO 521
金沢大学研究者情報 30507372
研究者番号 30507372

Matsubara, Kosuke

Search repository
書誌情報 European Radiology

巻 24, 号 8, p. 1794-1799, 発行日 2014-08-01
ISSN
収録物識別子タイプ ISSN
収録物識別子 0938-7994
NCID
収録物識別子タイプ NCID
収録物識別子 AA10824302
DOI
関連タイプ isVersionOf
識別子タイプ DOI
関連識別子 10.1007/s00330-014-3186-x
出版者
出版者 Springer Verlag
抄録
内容記述タイプ Abstract
内容記述 Objectives We invented a drape-like shield against scattered X-rays that can safely come into contact with medical equipment or people during fluoroscopically guided procedures. The shield can be easily removed from a C-arm unit using one hand. We evaluated the use of the novel removable shield during the endoscopic retrograde cholangiopancreatography (ERCP) procedure. Methods We measured the dose rate of scattered X-rays around endoscopists with and without this removable shield and surveyed the occupational doses to the ERCP staff. We also examined the endurance of the shield. Results The removable shield reduced the dose rate of scattered X-rays to one-tenth and reduced the monthly dose to an endoscopist by at least two-fifths. For 2.5 years, there was no damage to the shield and no loosening of the seam. The bonding of the hook-and-loop fasteners did not weaken, although the powerful double-sided tapes made especially for plastic did. Conclusions The removable shield can reduce radiation exposure to the ERCP staff and may contribute to reducing the exposure to the eye lenses of operators. It would also be possible to expand its use to other fluoroscopically guided procedures besides ERCP because it is a light, simple, and useful device. Key Points • We invented a shield that can be removed from C-arm units with one hand.• The removable shield reduces the dose rate of X-rays to one-tenth.• The removable shield reduces operator exposure by two-fifths.• The removable shield is durable, lasting for several years.• The drape-like removable shield is light, simple, and useful. © 2014 European Society of Radiology.
著者版フラグ
出版タイプ AM
出版タイプResource http://purl.org/coar/version/c_ab4af688f83e57aa
戻る
0
views
See details
Views

Versions

Ver.1 2023-07-28 00:52:24.534234
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3