A novel removable shield attached to C-arm units against scattered X-rays from a patient's side
http://hdl.handle.net/2297/39048
http://hdl.handle.net/2297/39048fd787266-6fd7-4fb2-a7b9-0d418ee1ae72
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-10-03 | |||||
タイトル | ||||||
タイトル | A novel removable shield attached to C-arm units against scattered X-rays from a patient's side | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
著者 |
Mori, Hiroshige
× Mori, Hiroshige× Koshida, Kichiro× Ishigamori, Osamu× Matsubara, Kosuke |
|||||
書誌情報 |
European Radiology 巻 24, 号 8, p. 1794-1799, 発行日 2014-08-01 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0938-7994 | |||||
NCID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AA10824302 | |||||
DOI | ||||||
関連タイプ | isVersionOf | |||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1007/s00330-014-3186-x | |||||
出版者 | ||||||
出版者 | Springer Verlag | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Objectives We invented a drape-like shield against scattered X-rays that can safely come into contact with medical equipment or people during fluoroscopically guided procedures. The shield can be easily removed from a C-arm unit using one hand. We evaluated the use of the novel removable shield during the endoscopic retrograde cholangiopancreatography (ERCP) procedure. Methods We measured the dose rate of scattered X-rays around endoscopists with and without this removable shield and surveyed the occupational doses to the ERCP staff. We also examined the endurance of the shield. Results The removable shield reduced the dose rate of scattered X-rays to one-tenth and reduced the monthly dose to an endoscopist by at least two-fifths. For 2.5 years, there was no damage to the shield and no loosening of the seam. The bonding of the hook-and-loop fasteners did not weaken, although the powerful double-sided tapes made especially for plastic did. Conclusions The removable shield can reduce radiation exposure to the ERCP staff and may contribute to reducing the exposure to the eye lenses of operators. It would also be possible to expand its use to other fluoroscopically guided procedures besides ERCP because it is a light, simple, and useful device. Key Points • We invented a shield that can be removed from C-arm units with one hand.• The removable shield reduces the dose rate of X-rays to one-tenth.• The removable shield reduces operator exposure by two-fifths.• The removable shield is durable, lasting for several years.• The drape-like removable shield is light, simple, and useful. © 2014 European Society of Radiology. | |||||