Application of a variable filter for presampled modulation transfer function analysis with the edge method

http://hdl.handle.net/2297/43456
http://hdl.handle.net/2297/43456
d1fe4341-7813-4674-9358-a169d5db3209
名前 / ファイル ライセンス アクション
ME-PR-HIGASHIDE-R-320.pdf ME-PR-HIGASHIDE-R-320.pdf (497.3 kB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-10-03
タイトル
タイトル Application of a variable filter for presampled modulation transfer function analysis with the edge method
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Higashide, Ryo

× Higashide, Ryo

WEKO 24476

Higashide, Ryo

Search repository
Ichikawa, Katsuhiro

× Ichikawa, Katsuhiro

WEKO 646
e-Rad 40402630
金沢大学研究者情報 40402630
研究者番号 40402630

Ichikawa, Katsuhiro

Search repository
Kunitomo, Hiroshi

× Kunitomo, Hiroshi

WEKO 24477

Kunitomo, Hiroshi

Search repository
Ohashi, Kazuya

× Ohashi, Kazuya

WEKO 24478

Ohashi, Kazuya

Search repository
書誌情報 Radiological Physics and Technology

巻 8, 号 2, p. 320-330, 発行日 2015-07-23
ISSN
収録物識別子タイプ ISSN
収録物識別子 1865-0333
NCID
収録物識別子タイプ NCID
収録物識別子 AA12236881
DOI
関連タイプ isVersionOf
識別子タイプ DOI
関連識別子 10.1007/s12194-015-0325-2
出版者
出版者 日本放射線技術学会 and 日本医学物理学会Japanese Society of Radiological Technology and Japan Society of Medical Physics / Springer-Verlag Tokyo
抄録
内容記述タイプ Abstract
内容記述 We devised a new noise filtering method to reduce the noise in the line spread function (LSF) for presampled modulation transfer function (MTF) analysis with the edge method. A filter was designed to reduce noise effectively using a position-dependent filter controlled by the boundary frequency b for low-pass filtering, which is calculated by 1/2d (d: distance from the LSF center). In this filtering process, strong filters with very low b can be applied to regions distant from the LSF center, and the region near the LSF center can be maintained simultaneously by a correspondingly high b. Presampled MTF accuracies derived by use of the proposed method and an edge spread function (ESF)-fitting method were compared by use of simulated ESFs with and without noise, resembling a computed radiography (CR) and an indirect-type flat panel detector (FPD), respectively. In addition, the edge images of clinical CR, indirect-type FPD, and direct-type FPD systems were examined. For a simulated ESF without noise, the calculated MTFs of the variable filtering method agreed precisely with the true MTFs. The excellent noise-reduction ability of the variable filter was demonstrated for all simulated noisy ESFs and those of three clinical systems. Although the ESF-fitting method provided excellent noise reduction only for the CR-like simulated ESF with noise, its noise elimination performance could not be demonstrated due to the lesser robustness of the fitting. © 2015, Japanese Society of Radiological Technology and Japan Society of Medical Physics.
内容記述
内容記述タイプ Other
内容記述 発行後1年より全文公開
権利
権利情報 The final publication is available at www.springerlink.com/article/10.1007%2Fs12194-015-0325-2
著者版フラグ
出版タイプ AM
出版タイプResource http://purl.org/coar/version/c_ab4af688f83e57aa
戻る
0
views
See details
Views

Versions

Ver.1 2023-07-28 00:50:44.391245
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3