Long-term stability of the Hounsfield unit to electron density calibration curve in cone-beam computed tomography images for adaptive radiotherapy treatment planning

http://hdl.handle.net/2297/44243
http://hdl.handle.net/2297/44243
3ef37a2e-baf9-4e76-8f7c-ccec7321aa6b
名前 / ファイル ライセンス アクション
ME-PR-TAKEMURA-A-410.pdf ME-PR-TAKEMURA-A-410.pdf (566.1 kB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-10-03
タイトル
タイトル Long-term stability of the Hounsfield unit to electron density calibration curve in cone-beam computed tomography images for adaptive radiotherapy treatment planning
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Takemura, Akihiro

× Takemura, Akihiro

WEKO 529
e-Rad 70313674
金沢大学研究者情報 70313674
研究者番号 70313674

Takemura, Akihiro

Search repository
Tanabe, Shogo

× Tanabe, Shogo

WEKO 24800

Tanabe, Shogo

Search repository
Tokai, Mei

× Tokai, Mei

WEKO 24801

Tokai, Mei

Search repository
Ueda, Shinichi

× Ueda, Shinichi

WEKO 24802

Ueda, Shinichi

Search repository
Noto, Kimiya

× Noto, Kimiya

WEKO 24803

Noto, Kimiya

Search repository
Isomura, Naoki

× Isomura, Naoki

WEKO 24804

Isomura, Naoki

Search repository
Kojima, Hironori

× Kojima, Hironori

WEKO 24805

Kojima, Hironori

Search repository
書誌情報 Journal of Radiotherapy in Practice

巻 14, 号 4, p. 410-417, 発行日 2015-06-11
ISSN
収録物識別子タイプ ISSN
収録物識別子 1460-3969
NCID
収録物識別子タイプ NCID
収録物識別子 AA11974173
DOI
関連タイプ isIdenticalTo
識別子タイプ DOI
関連識別子 10.1017/S1460396915000321
出版者
出版者 Cambridge University Press
抄録
内容記述タイプ Abstract
内容記述 Aim To use cone-beam computed tomography (CBCT) images for treatment planning, the Hounsfield unit (HU)-electron density (ED) calibration table for CBCT should be stable. The purpose of this study was to verify the stability of the HU values for the CBCT system over 1 year and to evaluate the effects of variation in HU-ED calibration curves on dose calculation. Materials and Methods A tissue characterisation phantom was scanned with the field of view (FOV) of size S (FOV-S) and FOV of size M (FOV-M) using the CBCT system once a month for 1 year. A single field treatment plan was constructed on digital phantom images to validate the dose distribution using mean HU-ED calibration curves and possible variations. Results HU values for each material rod over the observation period varied with trend. The HU value of the cortical bone rod decreased by about 100 HU for the FOV-S and by about 300 HU for the FOV-M. Possible variation in the HU-ED calibration curves produced a ≤17·9% dose difference in the dose maximum in the treatment plan. Conclusions The CBCT system should be calibrated periodically for consistent dose calculation. © Cambridge University Press 2015. Aim To use cone-beam computed tomography (CBCT) images for treatment planning, the Hounsfield unit (HU)-electron density (ED) calibration table for CBCT should be stable. The purpose of this study was to verify the stability of the HU values for the CBCT system over 1 year and to evaluate the effects of variation in HU-ED calibration curves on dose calculation. Materials and Methods A tissue characterisation phantom was scanned with the field of view (FOV) of size S (FOV-S) and FOV of size M (FOV-M) using the CBCT system once a month for 1 year. A single field treatment plan was constructed on digital phantom images to validate the dose distribution using mean HU-ED calibration curves and possible variations. Results HU values for each material rod over the observation period varied with trend. The HU value of the cortical bone rod decreased by about 100 HU for the FOV-S and by about 300 HU for the FOV-M. Possible variation in the HU-ED calibration curves produced a ≤17·9% dose difference in the dose maximum in the treatment plan. Conclusions The CBCT system should be calibrated periodically for consistent dose calculation. © Cambridge University Press 2015.
著者版フラグ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
戻る
0
views
See details
Views

Versions

Ver.1 2023-07-28 00:48:57.190821
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3