X-ray dose reduction using additional copper filtration for abdominal digital radiography: Evaluation using signal difference-to-noise ratio

http://hdl.handle.net/2297/47062
http://hdl.handle.net/2297/47062
b9bc49ec-e293-4e99-9f53-ada8c62f97f9
名前 / ファイル ライセンス アクション
ME-PR-KAWASHIMA-H-65.pdf ME-PR-KAWASHIMA-H-65.pdf (522.5 kB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-12-05
タイトル
タイトル X-ray dose reduction using additional copper filtration for abdominal digital radiography: Evaluation using signal difference-to-noise ratio
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Kawashima, Hiroki

× Kawashima, Hiroki

WEKO 13186
金沢大学研究者情報 70775577
研究者番号 70775577

Kawashima, Hiroki

Search repository
Ichikawa, Katsuhiro

× Ichikawa, Katsuhiro

WEKO 646
e-Rad 40402630
金沢大学研究者情報 40402630
研究者番号 40402630

Ichikawa, Katsuhiro

Search repository
Nagasou, Daisuke

× Nagasou, Daisuke

WEKO 68926

Nagasou, Daisuke

Search repository
Hattori, Masayuki

× Hattori, Masayuki

WEKO 68927

Hattori, Masayuki

Search repository
書誌情報 Physica Medica

巻 34, p. 65-71, 発行日 2017-02-01
ISSN
収録物識別子タイプ ISSN
収録物識別子 1120-1797
NCID
収録物識別子タイプ NCID
収録物識別子 AA11064306
DOI
関連タイプ isVersionOf
識別子タイプ DOI
関連識別子 10.1016/j.ejmp.2017.01.015
出版者
出版者 Associazione Italiana di Fisica Medica / Elsevier
抄録
内容記述タイプ Abstract
内容記述 Purpose: X-ray dose reduction using additional copper filters (Cu-filters) for abdominal general radiography was indicated in a report using a simulation study. We validated the dose reduction effects using a clinical digital radiography system equipped with an indirect-type CsI detector and an automatic Cu-filter insertion function. Methods: The image qualities were evaluated using signal difference-to-noise ratio (SDNR) for different radiation qualities with and without Cu-filters for a 20-cm acrylic phantom. Acrylic and bone equivalent material plates were used for contrast measurements. The dose reduction using Cu-filters was estimated from the ratios of the SDNR2 values. Results: For the same entrance surface dose (ESD), Cu-filters with 0.1- and 0.2-mm thicknesses increased the image quality as evaluated by SDNR2 and the estimated dose reduction without degrading the image quality. For the acrylic contrast, the dose reductions with the 0.1- and 0.2-mm-thick Cu-filters were approximately 30% and 44% at 70kV and 29% and 35% at 80kV, respectively. For the bone contrast, the reduction rates were slightly reduced. Conclusions: We validated the dose reduction capability of additional Cu-filters without degrading the image quality for abdominal radiography. The estimated entrance surface dose reductions of the Cu-filters were approximately 30-40% and 20-30% for the acrylic and bone contrasts, respectively, and effective dose reductions for acrylic were nearly half of those for ESD. At these reduced dose conditions, the current time product values needed to be increased by factors of 1.4 and 1.8 for the 0.1- and 0.2-mm-thick Cu-filters, respectively. © 2017 Associazione Italiana di Fisica Medica.
内容記述
内容記述タイプ Other
内容記述 Embargo Period 12 months
権利
権利情報 Copyright © Elsevier (CC-BY NC ND)
著者版フラグ
出版タイプ AM
出版タイプResource http://purl.org/coar/version/c_ab4af688f83e57aa
関連URI
識別子タイプ URI
関連識別子 http://www.elsevier.com/locate/issn/11201797
戻る
0
views
See details
Views

Versions

Ver.1 2023-07-27 17:46:51.599279
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3