Selective Peak Enhancement and Suppression in Conductometric Ion Chromatography and Its Application to the Determination of Trace Levels of Orthophosphate in Environmental Samples

http://hdl.handle.net/2297/44558
http://hdl.handle.net/2297/44558
33e06420-460a-4835-a704-577768d53090
名前 / ファイル ライセンス アクション
PH-PR-HAYAKAWA-K-25.pdf PH-PR-HAYAKAWA-K-25.pdf (300.0 kB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-10-04
タイトル
タイトル Selective Peak Enhancement and Suppression in Conductometric Ion Chromatography and Its Application to the Determination of Trace Levels of Orthophosphate in Environmental Samples
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Hayakawa, Kazuichi

× Hayakawa, Kazuichi

WEKO 35
e-Rad 40115267
金沢大学研究者情報 40115267
研究者番号 40115267

Hayakawa, Kazuichi

Search repository
Kato, Akio

× Kato, Akio

WEKO 27455

Kato, Akio

Search repository
Yamamoto, Atsushi

× Yamamoto, Atsushi

WEKO 27456

Yamamoto, Atsushi

Search repository
Miyazaki, Motoichi

× Miyazaki, Motoichi

WEKO 27264
e-Rad 50009164
研究者番号 50009164

Miyazaki, Motoichi

Search repository
書誌情報 Analytical Sciences

巻 8, 号 1, p. 25-29, 発行日 1992-01-01
ISSN
収録物識別子タイプ ISSN
収録物識別子 0910-6340
NCID
収録物識別子タイプ NCID
収録物識別子 AA10500785
DOI
関連タイプ isIdenticalTo
識別子タイプ DOI
関連識別子 10.2116/analsci.8.25
出版者
出版者 日本分析化学会 = The Japan Society for Analytical Chemistry
抄録
内容記述タイプ Abstract
内容記述 In non-suppressed conductometric ion chromatography, when a carboxylic acid eluent of low pH is used, the signal intensity (ΔHc) is proportional to the sum of the conductivity change caused by changes in the sample concentration (ΔCs) as well as conductivity changes caused by changes in the eluent concentration (ΔCe). This can be expressed as ΔHC=B(ΔCs)+B(ΔCe), where B is a constant. In this equation, ΔCs is constant for any sample species. However, ΔCe varies with changes in the sample capacity factors. This value is positive when the sample elutes before the system peak (k′s<k′e), and is negative when it elutes after the system peak (k′s>k′e). Therefore, the sample peak intensity (ΔHC) is enhanced prior to the system peak and is suppressed after it. These effects increase the closer the sample and the system peaks. The effect of this theory was proved by a determination of orthophosphate using a TSK gel IC-Anion-PW column and a 0.75mM phthalic acid eluent. The detection limit of this method for orthophosphate (1×10-11mol, with an S/N of three) was smaller (by a factor of 10-30) than those by previous conventional ion chromatographic methods. Also, the interference of chloride or nitrate, often observed in the previous methods, was removed by suppressing their peak intensities. When this method was applied for the determination of orthophosphate in environmental samples, other interfering compounds were effectively removed by a column-switching system attached to the ion chromatograph, and a trace level of the orthophosphate was determined.
権利
権利情報 Copyright © The Japan Society for Analytical Chemistry
著者版フラグ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
関連URI
識別子タイプ URI
関連識別子 https://www.jstage.jst.go.jp/browse/analsci
関連URI
識別子タイプ URI
関連識別子 http://www.jsac.jp/
戻る
0
views
See details
Views