  3. 1. 査読済論文

Inter-institution preference-based variability of ejection fraction and volumes using quantitative gated SPECT with Tc-99m tetrofosmin : A multi-center study involving 106 hospitals

http://hdl.handle.net/2297/2243
http://hdl.handle.net/2297/2243
541a4b26-68f6-4b21-9f96-89580d4f3ddd
名前 / ファイル ライセンス アクション
HO-PR-NAKAJIMA-K-2006QGSmulticenterRepository.pdf HO-PR-NAKAJIMA-K-2006QGSmulticenterRepository.pdf (878.8 kB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-10-05
タイトル
タイトル Inter-institution preference-based variability of ejection fraction and volumes using quantitative gated SPECT with Tc-99m tetrofosmin : A multi-center study involving 106 hospitals
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Nakajima, Kenichi

× Nakajima, Kenichi

WEKO 320
e-Rad 00167545
金沢大学研究者情報 00167545
研究者番号 00167545

Nakajima, Kenichi

Search repository
Nishimura, Tsunehiko

× Nishimura, Tsunehiko

WEKO 44999

Nishimura, Tsunehiko

Search repository
書誌情報 European Journal of Nuclear Medicine and Molecular Imaging

巻 33, 号 2, p. 127-133, 発行日 2006-02-01
ISSN
収録物識別子タイプ ISSN
収録物識別子 1619-7070
DOI
関連タイプ isVersionOf
識別子タイプ DOI
関連識別子 https://doi.org/10.1007/s00259-005-1916-7
出版者
出版者 Springer
抄録
内容記述タイプ Abstract
内容記述 Purpose Inter-institution reproducibility of gated SPECT quantification based on institutional preferences was evaluated. This sort of variability is crucial for a multicentre study when many hospitals are involved. Methods A total of 106 institutes participated in this study and were grouped according to their use of five workstation types. Fifteen sets of 99mTc-tetrofosmin gated projection images with normal ejection fraction (EF) (70%, group A, n=5), borderline low EF (50%, group B, n=5) and low EF with large perfusion defects (30%, group C, n=5) were prepared. The projection images were processed by QGS software in each institute based on its own routine settings. Based on 318 QGS results, the reproducibility of EF and volumes was analysed for each group and workstation. Results The reproducibility of EF was good in 14 of 15 cases, showing a standard deviation (SD) of <3.6%, and the coefficient of variance of the end-diastolic volume (EDV) was <9.3% in all cases. When the deviation from the average value was analysed, the difference between EF at each institute and the average EF of the workstation (dEF) showed an SD of 2.2–3.7% for each group. The ratio of the EDV divided by the average EDV (rEDV) showed an SD of 0.061–0.069 for each group. One case in group C that had a large anterior defect with low EF showed bimodal EF distribution in one of the five workstations. The SD of EF was workstation dependent, owing to the SPECT reconstruction conditions. Conclusion The reproducibility in EF and volumes within a workstation was good, even though the gated SPECT preferences varied. This reproducibility study supports the use of gated SPECT as a standard of ventricular function in multicentre studies.© Springer-Verlag 1999.
著者版フラグ
出版タイプ AM
出版タイプResource http://purl.org/coar/version/c_ab4af688f83e57aa
戻る
0
views
See details
Views

Versions

Ver.1 2023-07-27 12:15:00.202325
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3