WEKO3
インデックスリンク
アイテム
{"_buckets": {"deposit": "f6cedf8d-5c81-4ad8-997b-9da284a809de"}, "_deposit": {"created_by": 18, "id": "54395", "owners": [18], "pid": {"revision_id": 0, "type": "depid", "value": "54395"}, "status": "published"}, "_oai": {"id": "oai:kanazawa-u.repo.nii.ac.jp:00054395", "sets": ["2819"]}, "author_link": ["2545"], "item_9_biblio_info_8": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2020-03-05", "bibliographicIssueDateType": "Issued"}, "bibliographicPageStart": "1p.", "bibliographicVolumeNumber": "2014-04-01 – 2015-03-31", "bibliographic_titles": [{"bibliographic_title": "平成26(2014)年度 科学研究費補助金 奨励研究 研究概要"}, {"bibliographic_title": "2014 Research Project Summary", "bibliographic_titleLang": "en"}]}]}, "item_9_creator_33": {"attribute_name": "著者別表示", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Yoneyama, Hiroto"}], "nameIdentifiers": [{"nameIdentifier": "2545", "nameIdentifierScheme": "WEKO"}]}]}, "item_9_description_21": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "乳癌患者56症例のセンチネルリンパ節シンチグラフィにおいて、通常行っているエネルギーウィンドウの140keV±10%を5つのウィンドウ(4%ごと)に分割してプラナ像の撮像を行った。得られた5つの画像(128.8keV±2%, 134.4keV±2%, 140keV±2%, 145.6keV±2%, 151.2keV±2%)に対してセンチネルリンパ節とバックグラウンドにROIをとり、それぞれのエネルギーウィンドウごとにコントラストを算出した。また、5名の臨床医に読影を依頼して5段階で評価し(1 : 確実にない、2 : たぶんない、3 : 疑わしい、4 : たぶんある、5 : 確実にある)、平均が4以上の症例を全症例で割ったものをセンチネルリンパ節の検出率として評価した。145.6keV±2%のコントラストが最も優れていた(コントラスト : 0.743±0.20)。続いて140keV±2%(0.734±0.20), 151.2keV±2%(0.634±0.26), 134.4keV±2%(0.630±0.23)、の順でコントラストは低下し、最もコントラストが悪いのは128.8keV±2%(0.409±0.24)であった。140keV±10%から128.8keV±2%と134.4keV±2%を取り除いて145.6keV±6%にするとコントラストが0.635±0.22から0.730±0.20に有意に改善し(p=0.0063)、読影医の評価が4.74±0.58から4.94±0.20に有意に改善し(P\u003c0.0001)、検出率が94.8%から100%に改善した。センチネルリンパ節が注入部位に近い場合、低エネルギー領域は散乱線を多く含むため、145.6keV±6%で撮像するとよい。今回の検討のように5つのエネルギーウィンドウで撮像することにより撮像後に状況に応じて適切なエネルギーウィンドウを設定することができる。", "subitem_description_type": "Abstract"}]}, "item_9_description_22": {"attribute_name": "内容記述", "attribute_value_mlt": [{"subitem_description": "研究課題/領域番号:26931049, 研究期間(年度):2014-04-01 – 2015-03-31", "subitem_description_type": "Other"}, {"subitem_description": "出典:研究課題「センチネルリンパ節シンチグラフィにおけるオフセットエネルギーウィンドウ撮像の検討」課題番号26931049\n(KAKEN:科学研究費助成事業データベース(国立情報学研究所)) \n(https://kaken.nii.ac.jp/grant/KAKENHI-PROJECT-26931049/)を加工して作成", "subitem_description_type": "Other"}]}, "item_9_description_5": {"attribute_name": "提供者所属", "attribute_value_mlt": [{"subitem_description": "金沢大学附属病院", "subitem_description_type": "Other"}]}, "item_9_identifier_registration": {"attribute_name": "ID登録", "attribute_value_mlt": [{"subitem_identifier_reg_text": "10.24517/00060671", "subitem_identifier_reg_type": "JaLC"}]}, "item_9_relation_28": {"attribute_name": "関連URI", "attribute_value_mlt": [{"subitem_relation_name": [{"subitem_relation_name_text": "https://kaken.nii.ac.jp/grant/KAKENHI-PROJECT-26931049/"}], "subitem_relation_type_id": {"subitem_relation_type_id_text": "https://kaken.nii.ac.jp/grant/KAKENHI-PROJECT-26931049/", "subitem_relation_type_select": "URI"}}]}, "item_9_version_type_25": {"attribute_name": "著者版フラグ", "attribute_value_mlt": [{"subitem_version_resource": "http://purl.org/coar/version/c_ab4af688f83e57aa", "subitem_version_type": "AM"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "米山, 寛人"}], "nameIdentifiers": [{"nameIdentifier": "2545", "nameIdentifierScheme": "WEKO"}]}]}, "item_files": {"attribute_name": "ファイル情報", "attribute_type": "file", "attribute_value_mlt": [{"accessrole": "open_date", "date": [{"dateType": "Available", "dateValue": "2021-06-21"}], "displaytype": "detail", "download_preview_message": "", "file_order": 0, "filename": "HO-PR-YONEYAMA-H-kaken 2020-1p.pdf", "filesize": [{"value": "122.3 kB"}], "format": "application/pdf", "future_date_message": "", "is_thumbnail": false, "licensetype": "license_11", "mimetype": "application/pdf", "size": 122300.0, "url": {"label": "HO-PR-YONEYAMA-H-kaken 2020-1p.pdf", "url": "https://kanazawa-u.repo.nii.ac.jp/record/54395/files/HO-PR-YONEYAMA-H-kaken 2020-1p.pdf"}, "version_id": "86b04073-cee6-4acf-aa68-3ae43d396890"}]}, "item_keyword": {"attribute_name": "キーワード", "attribute_value_mlt": [{"subitem_subject": "センチネルリンパ節シンチグラフィ", "subitem_subject_scheme": "Other"}, {"subitem_subject": "散乱線", "subitem_subject_scheme": "Other"}, {"subitem_subject": "コンプトン散乱", "subitem_subject_scheme": "Other"}, {"subitem_subject": "Clinical medicine", "subitem_subject_language": "en", "subitem_subject_scheme": "Other"}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "jpn"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "research report", "resourceuri": "http://purl.org/coar/resource_type/c_18ws"}]}, "item_title": "センチネルリンパ節シンチグラフィにおけるオフセットエネルギーウィンドウ撮像の検討", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "センチネルリンパ節シンチグラフィにおけるオフセットエネルギーウィンドウ撮像の検討"}]}, "item_type_id": "9", "owner": "18", "path": ["2819"], "permalink_uri": "https://doi.org/10.24517/00060671", "pubdate": {"attribute_name": "公開日", "attribute_value": "2021-06-21"}, "publish_date": "2021-06-21", "publish_status": "0", "recid": "54395", "relation": {}, "relation_version_is_last": true, "title": ["センチネルリンパ節シンチグラフィにおけるオフセットエネルギーウィンドウ撮像の検討"], "weko_shared_id": -1}
センチネルリンパ節シンチグラフィにおけるオフセットエネルギーウィンドウ撮像の検討
https://doi.org/10.24517/00060671
https://doi.org/10.24517/0006067125adff22-6529-4084-a7ec-a150a1c37cab
名前 / ファイル | ライセンス | アクション |
---|---|---|
HO-PR-YONEYAMA-H-kaken 2020-1p.pdf (122.3 kB)
|
Item type | 報告書 / Research Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2021-06-21 | |||||
タイトル | ||||||
タイトル | センチネルリンパ節シンチグラフィにおけるオフセットエネルギーウィンドウ撮像の検討 | |||||
言語 | ||||||
言語 | jpn | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_18ws | |||||
資源タイプ | research report | |||||
ID登録 | ||||||
ID登録 | 10.24517/00060671 | |||||
ID登録タイプ | JaLC | |||||
著者別表示 |
Yoneyama, Hiroto
× Yoneyama, Hiroto |
|||||
提供者所属 | ||||||
内容記述タイプ | Other | |||||
内容記述 | 金沢大学附属病院 | |||||
書誌情報 |
平成26(2014)年度 科学研究費補助金 奨励研究 研究概要 en : 2014 Research Project Summary 巻 2014-04-01 – 2015-03-31, p. 1p., 発行日 2020-03-05 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | 乳癌患者56症例のセンチネルリンパ節シンチグラフィにおいて、通常行っているエネルギーウィンドウの140keV±10%を5つのウィンドウ(4%ごと)に分割してプラナ像の撮像を行った。得られた5つの画像(128.8keV±2%, 134.4keV±2%, 140keV±2%, 145.6keV±2%, 151.2keV±2%)に対してセンチネルリンパ節とバックグラウンドにROIをとり、それぞれのエネルギーウィンドウごとにコントラストを算出した。また、5名の臨床医に読影を依頼して5段階で評価し(1 : 確実にない、2 : たぶんない、3 : 疑わしい、4 : たぶんある、5 : 確実にある)、平均が4以上の症例を全症例で割ったものをセンチネルリンパ節の検出率として評価した。145.6keV±2%のコントラストが最も優れていた(コントラスト : 0.743±0.20)。続いて140keV±2%(0.734±0.20), 151.2keV±2%(0.634±0.26), 134.4keV±2%(0.630±0.23)、の順でコントラストは低下し、最もコントラストが悪いのは128.8keV±2%(0.409±0.24)であった。140keV±10%から128.8keV±2%と134.4keV±2%を取り除いて145.6keV±6%にするとコントラストが0.635±0.22から0.730±0.20に有意に改善し(p=0.0063)、読影医の評価が4.74±0.58から4.94±0.20に有意に改善し(P<0.0001)、検出率が94.8%から100%に改善した。センチネルリンパ節が注入部位に近い場合、低エネルギー領域は散乱線を多く含むため、145.6keV±6%で撮像するとよい。今回の検討のように5つのエネルギーウィンドウで撮像することにより撮像後に状況に応じて適切なエネルギーウィンドウを設定することができる。 | |||||
内容記述 | ||||||
内容記述タイプ | Other | |||||
内容記述 | 研究課題/領域番号:26931049, 研究期間(年度):2014-04-01 – 2015-03-31 | |||||
内容記述 | ||||||
内容記述タイプ | Other | |||||
内容記述 | 出典:研究課題「センチネルリンパ節シンチグラフィにおけるオフセットエネルギーウィンドウ撮像の検討」課題番号26931049 (KAKEN:科学研究費助成事業データベース(国立情報学研究所)) (https://kaken.nii.ac.jp/grant/KAKENHI-PROJECT-26931049/)を加工して作成 |
|||||
著者版フラグ | ||||||
出版タイプ | AM | |||||
出版タイプResource | http://purl.org/coar/version/c_ab4af688f83e57aa | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://kaken.nii.ac.jp/grant/KAKENHI-PROJECT-26931049/ | |||||
関連名称 | https://kaken.nii.ac.jp/grant/KAKENHI-PROJECT-26931049/ |