Noninvasive measurement of instantaneous, radial artery blood pressure

http://hdl.handle.net/2297/9537
http://hdl.handle.net/2297/9537
237b2fe1-e88a-47d4-94ff-92349ba099b1
名前 / ファイル ライセンス アクション
preparation.txt preparation.txt (71 Bytes)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-10-03
タイトル
タイトル Noninvasive measurement of instantaneous, radial artery blood pressure
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Tanaka, S.

× Tanaka, S.

WEKO 10447

Tanaka, S.

Search repository
Gao, Shumei

× Gao, Shumei

WEKO 10448

Gao, Shumei

Search repository
Nogawa, M.

× Nogawa, M.

WEKO 10449

Nogawa, M.

Search repository
Yamakoshi, Kenichi

× Yamakoshi, Kenichi

WEKO 42
e-Rad 40014310
研究者番号 40014310

Yamakoshi, Kenichi

Search repository
書誌情報 Engineering in Medicine and Biology Magazine, IEEE

巻 24, 号 4, p. 32-37, 発行日 2005-01-01
DOI
関連タイプ isIdenticalTo
識別子タイプ DOI
関連識別子 10.1109/MEMB.2005.1463393
出版者
出版者 IEEE
抄録
内容記述タイプ Abstract
内容記述 A new prototype system for the noninvasive measurement of instantaneous blood pressure (BP) in the human radial artery has been developed. The system is based on the volume-compensation method, but several components were newly designed, namely, a disk-type cuff for local pressurization and a nozzle-flapper-type electro-pneumatic converter (EPC) for the cuff-pressure control. Results of experiments indicated that, using the disk-type cuff, the radial artery could be completely compressed, and the newly devised EPC had sufficient frequency response for BP measurement in humans. It was also found that the prototype system comprising these components was capable of noninvasively measuring instantaneous BP in the human radial artery, not only under rest conditions but also in a stressful condition, such as during the cold pressor test. In conclusion, the newly developed local pressurization technique appears promising as a useful and helpful means for instantaneous BP monitoring over longer periods of time than have previously been possible. Using this method, it is possible to construct an instrument that is more compact and more comfortable for the subject to use than a conventional commercial instrument. Finally, it has been reported that it may be possible to monitor cardiac output continuously by BP waveform analysis. Thus, continuous cardiac output monitoring might be another potential and important application of the newly designed instrument.
著者版フラグ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
親アイテム(IsPartOf)
関連タイプ isPartOf
識別子タイプ URI
関連識別子 http://hdl.handle.net/2297/9285
戻る
0
views
See details
Views

Versions

Ver.1 2023-07-28 02:20:18.678880
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3