Development of a near-infrared spectroscopic system for monitoring urine glucose level for the use of long-term home healthcare

http://hdl.handle.net/2297/27505
http://hdl.handle.net/2297/27505
2c0f6333-735a-475e-b91e-e67b24fb922d
名前 / ファイル ライセンス アクション
TE-PR-TANAKA-S.pdf TE-PR-TANAKA-S.pdf (189.8 kB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-10-03
タイトル
タイトル Development of a near-infrared spectroscopic system for monitoring urine glucose level for the use of long-term home healthcare
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Tanaka, Shinobu

× Tanaka, Shinobu

WEKO 78
e-Rad 40242218
金沢大学研究者情報 40242218
研究者番号 40242218

Tanaka, Shinobu

Search repository
Hayakawa, Yuuto

× Hayakawa, Yuuto

WEKO 11647

Hayakawa, Yuuto

Search repository
Ogawa, Mitsuhiro

× Ogawa, Mitsuhiro

WEKO 11648

Ogawa, Mitsuhiro

Search repository
Yamakoshi, Ken-ichi

× Yamakoshi, Ken-ichi

WEKO 42
e-Rad 40014310
研究者番号 40014310

Yamakoshi, Ken-ichi

Search repository
書誌情報 Proceedings of SPIE - The International Society for Optical Engineering

巻 7544, 発行日 2010-01-01
ISSN
収録物識別子タイプ ISSN
収録物識別子 0277-786X
NCID
収録物識別子タイプ NCID
収録物識別子 AA10619755
DOI
関連タイプ isIdenticalTo
識別子タイプ DOI
関連識別子 10.1117/12.886007
出版者
出版者 Society of Photo-optical Instrumentation Engineers (SPIE)
抄録
内容記述タイプ Abstract
内容記述 We have been developing a new technique for measuring urine glucose concentration using near infrared spectroscopy (NIRS) in conjunction with the Partial Least Square (PLS) method. In the previous study, we reported some results of preliminary experiments for assessing feasibility of this method using a FT-IR spectrometer. In this study, considering practicability of the system, a flow-through cell with the optical path length of 10 mm was newly introduced. Accuracy of the system was verified by the preliminary experiments using urine samples. From the results obtained, it was clearly demonstrated that the present method had a capability of predicting individual urine glucose level with reasonable accuracy (the minimum value of standard error of prediction: SEP = 22.3 mg/dl) and appeared to be a useful means for long-term home health care. However, mean value of SEP obtained by the urine samples from ten subjects was not satisfactorily low (53.7 mg/dl). For improving the accuracy, (1) mechanical stability of the optical system should be improved, (2) the method for normalizing the spectrum should be reconsidered, and (3) the number of subject should be increased.
著者版フラグ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
戻る
0
views
See details
Views

Versions

Ver.1 2023-07-28 02:10:12.386450
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3