Temporal resolution measurement of 128-slice dual source and 320-row area detector computed tomography scanners in helical acquisition mode using the impulse method

http://hdl.handle.net/2297/45574
http://hdl.handle.net/2297/45574
fd4436f4-edd1-4c04-bbb5-bdcc5585b14b
名前 / ファイル ライセンス アクション
ME-PR-ICHIKAWA-K-625.pdf ME-PR-ICHIKAWA-K-625.pdf (628.1 kB)
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-10-03
タイトル
タイトル Temporal resolution measurement of 128-slice dual source and 320-row area detector computed tomography scanners in helical acquisition mode using the impulse method
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Hara, Takanori

× Hara, Takanori

WEKO 24446

Hara, Takanori

Search repository
Urikura, Atsushi

× Urikura, Atsushi

WEKO 24447

Urikura, Atsushi

Search repository
Ichikawa, Katsuhiro

× Ichikawa, Katsuhiro

WEKO 646
e-Rad 40402630
金沢大学研究者情報 40402630
研究者番号 40402630

Ichikawa, Katsuhiro

Search repository
Hoshino, Takashi

× Hoshino, Takashi

WEKO 24448

Hoshino, Takashi

Search repository
Nishimaru, Eiji

× Nishimaru, Eiji

WEKO 24449

Nishimaru, Eiji

Search repository
Niwa, Shinji

× Niwa, Shinji

WEKO 24450

Niwa, Shinji

Search repository
書誌情報 Physica Medica

巻 32, 号 4, p. 625-630, 発行日 2016-04-01
ISSN
収録物識別子タイプ ISSN
収録物識別子 1120-1797
NCID
収録物識別子タイプ NCID
収録物識別子 AA11064306
DOI
関連タイプ isVersionOf
識別子タイプ DOI
関連識別子 10.1016/j.ejmp.2016.02.015
出版者
出版者 Associazione Italiana di Fisica Medica / Elsevier
抄録
内容記述タイプ Abstract
内容記述 Purpose: To analyse the temporal resolution (TR) of modern computed tomography (CT) scanners using the impulse method, and assess the actual maximum TR at respective helical acquisition modes. Methods: To assess the actual TR of helical acquisition modes of a 128-slice dual source CT (DSCT) scanner and a 320-row area detector CT (ADCT) scanner, we assessed the TRs of various acquisition combinations of a pitch factor (P) and gantry rotation time (R). Results: The TR of the helical acquisition modes for the 128-slice DSCT scanner continuously improved with a shorter gantry rotation time and greater pitch factor. However, for the 320-row ADCT scanner, the TR with a pitch factor of <1.0 was almost equal to the gantry rotation time, whereas with pitch factor of >1.0, it was approximately one half of the gantry rotation time. The maximum TR values of single- and dual-source helical acquisition modes for the 128-slice DSCT scanner were 0.138 (R/. P = 0.285/1.5) and 0.074. s (R/. P = 0.285/3.2), and the maximum TR values of the 64. ×. 0.5- and 160. ×. 0.5-mm detector configurations of the helical acquisition modes for the 320-row ADCT scanner were 0.120 (R/. P = 0.275/1.375) and 0.195. s (R/. P = 0.3/0.6), respectively. Conclusion: Because the TR of a CT scanner is not accurately depicted in the specifications of the individual scanner, appropriate acquisition conditions should be determined based on the actual TR measurement. © 2016 Associazione Italiana di Fisica Medica.
権利
権利情報 Copyright © Elsevier (CC-BY NC ND)
著者版フラグ
出版タイプ AM
出版タイプResource http://purl.org/coar/version/c_ab4af688f83e57aa
関連URI
識別子タイプ URI
関連識別子 http://www.elsevier.com/locate/issn/11201797
戻る
0
views
See details
Views

Versions

Ver.1 2023-07-28 00:50:51.391749
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3