WEKO3
インデックスリンク
アイテム
圧—容積曲線変曲点に着目した容積振動型収縮期・拡張期血圧決定法の提案
http://hdl.handle.net/2297/35375
http://hdl.handle.net/2297/353754c3906c7-ed85-4877-8fb1-524271fb137f
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-10-03 | |||||
タイトル | ||||||
タイトル | 圧—容積曲線変曲点に着目した容積振動型収縮期・拡張期血圧決定法の提案 | |||||
言語 | ||||||
言語 | jpn | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
著者 |
野川, 雅道
× 野川, 雅道× 山越, 健弘× 松村, 健太× 田中, 志信× 小川, 充洋× 本井, 幸介× 山越, 憲一 |
|||||
書誌情報 |
生体医工学 = Transactions of Japanese Society for Medical and Biological Engineering : BME 巻 49, 号 6, p. 968-976, 発行日 2011-01-01 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 1347-443X | |||||
NCID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AA12119469 | |||||
DOI | ||||||
関連タイプ | isIdenticalTo | |||||
識別子タイプ | DOI | |||||
関連識別子 | 10.11239/jsmbe.49.968 | |||||
出版者 | ||||||
出版者 | 日本生体医工学会 = Japanese Society for Medical and Biological Engineering | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | A new method is presented for the indirect measurement of systolic (SBP) and diastolic blood pressure (DBP) in the radial artery. We use the volume-oscillometric method with a wrist cuff and then, utilizing the photo-plethysmograph (PPG) signal together with cuff pressure measurements, we derive pressure-volume curves (PVC). The principle of this method, named “PVC inflection-point method”, is to identify the indirect SBP and DBP using the inflection-points of envelope curves overlaying respectively the lower and upper points of the photo-plethysmographic pulsation signals superimposed on an arterial volume signal during the volume-oscillometric procedure. The accuracy of the method was evaluated by comparison with measurements from arterial catheterization in two healthy subjects. During the test procedure, a prototype PPG sensor for the PVC inflection-point method was attached to the left wrist, and arterial catheterization of the right radial artery for was performed to obtain simultaneous blood pressure recordings. The Bland-Altman plots for the comparative analysis showed good agreements between the two methods for both SBP (+2.3±7.7 (mean±S.D.) mmHg) and DBP (-1.8±6.4mmHg) measurement. We conclude that the proposed PVC inflection-point method appears promising to determine indirect SBP and DBP values with appropriate accuracy for convenient and practical use. | |||||
権利 | ||||||
権利情報 | Copyright © 日本生体医工学会 Japanese Society for Medical and Biological Engineering | |||||
著者版フラグ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | http://jsmbe.org/ | |||||
関連URI | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://www.jstage.jst.go.jp/browse/jsmbe/-char/ja/ |